The Agilent/Keysight 35677A is a 50-ohm S-parameter test set that extends the measurement capabilities of the 3577A Network Analyzer to characterize one- and two-port devices from 100 kHz to 200 MHz. It delivers full four-parameter S-parameter measurements (S11, S21, S12, S22) with directivity exceeding 40 dB, enabling precise reflection and transmission analysis in both forward and reverse directions.
Technical Specifications
Frequency Range: 100 kHz to 200 MHz
Test Port Impedance: 50 ohms
Directivity: >40 dB
S-Parameter Measurement Capabilities: S11, S21, S12, S22
Frequency Response Accuracy:
• Transmission (S21, S12): ±1 dB, ±5°
• Reflection (S11, S22): ±1 dB, ±5°
Port Specifications:
• Test Ports 1, 2 match: >26 dB
• Input port return loss: >20 dB
• Test ports open/short magnitude: <±0.75 dB
• Test ports open/short phase: <±5°
Insertion Loss:
• RF input to test port: typically 13 dB
• RF input to output ports A, B, R: typically 19 dB
Input Power Limits:
• Maximum operating level: +25 dBm or ±30 Vdc
• Damage level: ±27 dBm or ±30 Vdc
• Test port damage level: +27 dBm or ±30 Vdc
Connectors: Type N female, 50 ohm (all ports)
– Key Features
• Sweep modes: linear, alternate, CW, log frequency, log amplitude
• Frequency sweep time: 100 ms/span to 6553 sec/span
• Amplitude sweep time: 1 ms/step to 16 s/step
• Continuous, single, and manual sweep modes with free run, immediate, line, and external triggering
• Three-term vector error correction and vector normalization
• Vector noise averaging with selectable resolution bandwidth filtering
• Direct group delay measurement via phase aperture division
– Compatibility & Integration
Designed exclusively for integration with the Agilent/Keysight 3577A Network Analyzer.