Call +1 ‪(484) 841-9341‬

NIST TRACEABILITY 

NIST-traceable calibration certificate available with data

Service Support

Service plans available for the products’ durability

FAST REPAIRS

1–3 day turnaround, 15-day return on repair policy

Software Updates 

Staying up-to-date with latest software updates.

Agilent/Keysight N1090A

SKU: N1090ACategories: Sampling Oscilloscopes
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

Agilent / Keysight Sampling Oscilloscopes. Wavelength Range: 750 nm to 1650 nm.

Request CalibrationRequest Repair
Product Starting From:

$15,900.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote N1090A
Equipment info

The Agilent/Keysight N1090A is a compact sampling oscilloscope module for optical and electrical waveform analysis in high-speed digital and optical signal characterization. Operating as an integrated module within the Keysight 86100 DCA-X mainframe, it delivers optical transmitter compliance testing and waveform acquisition through PC-based control via the N1010A FlexDCA software over USB 2.0 or 3.0. The module supports data rates from 1 Gb/s to 10 Gb/s across wavelengths from 750 nm to 1650 nm with factory-calibrated optical-to-electrical conversion at 850 nm, 1310 nm, and 1550 nm (±20 nm each).

Technical Specifications

• Wavelength Range: 750 nm to 1650 nm (multimode or single-mode)
• Factory Calibrated Wavelengths: 850 nm (±20 nm), 1310 nm (±20 nm), 1550 nm (±20 nm)
• Data Rate Support: 1 Gb/s to 10 Gb/s
• Optical Noise Floor: As low as 1 µW
• Operating Temperature: 10 °C to +40 °C
• Non-operating Temperature: –40 °C to +70 °C
• Maximum Altitude: 4,600 meters (15,000 ft)
• Maximum Humidity: 80% (up to 31 °C)
• Power Consumption: 290 Watts maximum
• Dimensions (without connectors/feet): 88.26 mm H × 207.40 mm W × 485 mm D
• Dimensions (with connectors/feet): 103.31 mm H × 219.56 mm W × 517.80 mm D
• Weight: 6.20 kg
• Communication: SCPI over LAN (VXI-11 or HiSlip)
• Triggering: Front Panel or Free Run (pattern lock not available) – Key Features • High-sensitivity calibrated optical reference receivers with industry-standard compliance tolerances
• Reference receiver filter options for GPON/Gb Ethernet (1.244–1.250 Gb/s), OC-48/STM-16 (2.488–2.500 Gb/s), 10 Gb Ethernet LX-4 (3.125 Gb/s), and broadband 1.25–11.3 Gb/s configurations
• SIRC (System Impulse Response Correction) filters enabling up to 50% bandwidth improvement
• All optical channels include reference receivers at 20.6, 25.78, 26.56, 27.95, and 28.05 Gb/s
• User interface consistent with modern FlexDCA on 86100D and N1000A platforms – Typical Applications • Optical transmitter compliance testing
• Optical and electrical waveform characterization
• Research and development environments
• High-volume manufacturing test floors – Compatibility & Integration • Keysight 86100 DCA-X mainframe acquisition system
• N1010A FlexDCA software (host PC control)
• USB 2.0/3.0 and LAN connectivity

MPN

N1090A

Wavelength Range

750 nm to 1650 nm

Brand Name

Agilent / Keysight