The Boonton 4230 is a digital signal processor-based RF power meter, available in single or dual-channel configurations. It is designed for accurate and reliable measurement of RF power levels across a wide range, from -70 dBm to +44 dBm, with specific frequency and power ranges dependent on the chosen sensor. This instrument is compatible with a broad spectrum of Boonton 51000 series power sensors, including diode, thermocouple, and waveguide types, enabling measurements from 10 kHz up to 100 GHz. The Boonton 4230 features selectable ranging and filtering, an automatic zeroing function, and a built-in 50 MHz precision calibrator. It offers a measurement speed of up to 200 readings per second for single-channel operation and 100 readings per second for dual-channel operation. The unit includes a 20-character, 4-line backlit LCD display that can show simultaneous readings for dual channels, along with bar graph indicators. For integration into automated test equipment (ATE) systems, the Boonton 4230 is equipped with a standard IEEE-488 interface and an optional RS-232 serial interface, with emulation capabilities for HP437, HP438, and Boonton 4220A.
### Technical Specifications
* **Model:** Boonton 4230
* **Type:** Digital Signal Processor (DSP) based RF Power Meter
* **Channels:** Single or Dual
* **Measurement Range:** -70 dBm to +44 dBm (sensor dependent)
* **Frequency Range:** 10 kHz to 100 GHz (sensor dependent)
* **Dynamic Range:** Up to 90 dB with diode sensors; 50 dB with thermocouple sensors
* **Measurement Speed:**
* Single Channel: > 200 readings/second
* Dual Channel: 100 readings/second
* **Display:**
* 20-character, 4-line backlit LCD
* Simultaneous display of dual channels with numeric and bar graph information
* 5-digit resolution
* **Ranging:** Selectable (seven ranges or autoranging)
* **Filtering:** Selectable filter times (up to 20 seconds in 50-millisecond increments)
* **Zeroing:** Automatic zeroing (nulling of offsets) independently on each range
* **Calibrator:** Built-in 50 MHz precision calibrator
* **Power Sensors:** Compatible with Boonton 51000 series diode, thermocouple, and waveguide sensors
* Sensor data adapters allow automatic download of sensor calibration data
* Instrument non-volatile memory can store calibration data for up to four sensors
* **Emulation:** HP437, HP438, and Boonton 4220A
### Electrical Specifications
* **Power Consumption:** 24 VA maximum
* **AC Input:** 100, 120, 220, or 240 VAC (±10%), 50 to 60 Hz or 400 Hz
* **Calibrator Output:** 0 dBm, 50 MHz
* Level Accuracy: ±0.7% (25°C) for 90 days; ±0.9% RSS, 1.2% worst case (0° to +55°C) for 1 year
### Interfaces
* **Standard:** IEEE-488 (GPIB)
* **Optional:** RS-232 serial interface
* Selectable baud rates, data, stop, and parity bits are programmable
### Physical Specifications
* **Dimensions:** 8.26 in. (21.0 cm) wide, 3.48 in. (8.9 cm) high, 13.5 in. (34.3 cm) deep
* **Weight:** 7 lbs (3.2 kg)
* **Operating Temperature:** 0° to +55 °C
### Outputs
* **Chart Recorder Output:** 0 to 10 Volt DC output, proportional to measurement values
### Compatibility
* **Power Sensors:** Compatible with all Boonton CW diode, thermocouple, and waveguide sensors
* **ATE Systems:** Integrates with ATE systems via IEEE-488 or optional RS-232 interfaces
* **Software:** Available LabVIEW drivers for data analysis and reporting
Why Choose Aumictech for Boonton 4230?
At Aumictech, we specialize in supplying high-end RF power measurement equipment. Whether you need accurate power readings across a broad frequency spectrum, reliable performance in demanding test environments, or seamless integration into your existing ATE systems, our team ensures the Boonton 4230 delivers maximum performance for your application.