Call +1 ‪(484) 841-9341‬

NIST TRACEABILITY 

NIST-traceable calibration certificate available with data

Service Support

Service plans available for the products’ durability

FAST REPAIRS

1–3 day turnaround, 15-day return on repair policy

Software Updates 

Staying up-to-date with latest software updates.

California Instruments 3000LX AC Power Source

SKU: 3000LXCategories: Other Test Equipment
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The California Instruments 3000LX is a high-performance AC power source designed for a wide range of testing and power conditioning applications. It delivers clean and reliable AC power with programmable voltage, frequency, and current. Ideal for simulating worldwide AC power conditions and testing electronic equipment.

Request CalibrationRequest Repair
Product Starting From:

$5,200.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote 3000LX
Equipment info

The California Instruments 3000LX is a high-performance AC power source and analyzer designed to simulate a wide range of global AC power conditions and test electronic equipment. This microprocessor-controlled instrument offers programmable voltage, frequency, and current, with a focus on delivering clean, reliable power for demanding applications. Its capabilities extend to simulating transient conditions and analyzing power quality, making it suitable for advanced testing scenarios.

### Technical Specifications

**General:**

* **Manufacturer:** California Instruments (Ametek Programmable Power)
* **Model:** 3000LX
* **Series:** Lx Series
* **Power Output:** 3 kVA (single phase), 1000 VA (three phase) per chassis. Multi-chassis configurations can achieve up to 18 kVA.
* **Phase Modes:** Single-phase and Three-phase operation. The 3000Lx model can be operated from a single-phase AC input.
* **Input Voltage:** Configurable, with options such as 208-230 V L-L and 400 V L-L. Input voltage specifications are for Line to Line (L-L) for three-phase, delta or wye configurations.
* **Output Voltage Ranges:** Programmable, with options for 0-150V and 0-300V AC. Alternative voltage range pairs are available with -HV and -EHV options.
* **Frequency Range:** Programmable down to 17 Hz with a lower voltage limit. Standard frequency range is typically up to 5 kHz, with some sources indicating up to 819.1 Hz or 1000 Hz for specific models or options.
* **Power Factor:** 0 to unity at full output VA. Typical power factor is 0.6.
* **Efficiency:** 75% typical.

**Output Characteristics:**

* **Voltage Regulation (ALC on):** Less than 0.1% Full Scale (FS).
* **Line Regulation:** Less than 0.02% for a 10% line change.
* **Programming Accuracy:**
* Voltage (rms): ±(0.05% + 0.25) V from 5.0 V to Full Scale.
* Frequency: ±0.025% (45 Hz – 819.1 Hz), ±0.07% (> 819.1 Hz).
* Phase: ±1° (45-100 Hz).
* **Output Noise:** 100mV rms typical (20 kHz to 1 MHz). 250mV rms typical for -HF option.
* **Harmonic Distortion:** Less than 1% (at full scale voltage, full resistive load).
* **Current Limit:** Programmable from 0 Amps to the maximum current for the selected range.
* **Peak Current:**
* 3000Lx: 5.7 x (Irms @ full scale voltage)
* 4500Lx: 3.8 x (Irms @ full scale voltage)
* 6000Lx: 3 x (Irms @ full scale voltage)
* **Inrush Current (Per phase):**
* @ 180-254 V: 50 A peak.
* @ 360-440 V: 83 A peak.

**Waveform Capabilities:**

* **Standard Waveforms:** Sinewave, Squarewave, Clipped Sinewave.
* **Arbitrary Waveform Generation:** User-defined waveforms can be downloaded over the bus. Capable of generating harmonic waveforms using DSP technology.
* **Transient Programming:** Ability to simulate AC line disturbances and transient steps.

**Measurements and Analysis:**

* **Built-in Power Analyzer:** Performs voltage and load current harmonic analysis and waveform acquisition. Capable of handling applications that would traditionally require multiple instruments.
* **Harmonics Measurement:** Measures voltage and current amplitude and phase harmonics up to the 50th harmonic (for fundamental frequencies up to 250 Hz). Measures Total Harmonic Distortion (THD) for both voltage and current. Harmonic analysis data can be displayed on the front panel or via PC GUI.
* **Waveform Acquisition:** Real-time digitization of voltage and current waveforms using a 4K sample buffer.
* **Parameters Measured:** RMS voltage, RMS current, real power, apparent power, harmonic distortion, frequency, phase angle.
* **Frequency Measurement Accuracy:** Valid for output > 30 Vrms (or > 20 Vrms in some sources).

**Interfaces:**

* **Standard:** IEEE-488 (GPIB), USB, and RS232C interfaces are included for ATE system integration.
* **RS232C:** Bi-directional serial interface with 9-pin D-shell connector. Supports handshake (CTS, RTS), 7 or 8 data bits, 2 stop bits, and baud rates from 9600 to 115200. Supplied with RS232C cable. SCPI command language is supported.
* **Optional:** Ethernet LAN interface.
* **Compatibility:** Functionally and bus-compatible with Agilent HP6834B and California Instruments iL Series AC Sources.

**Physical:**

* **Dimensions (per chassis):**
* Height: 10.5” (267 mm)
* Width: 19” (483 mm)
* Depth: 23.7” (602 mm) (includes rear panel connectors)
* **Weight (per chassis):**
* Net: 193 lbs / 87.7 Kg
* Shipping: 280 lbs / 127.3 Kg
* For multi-chassis configurations, multiply by the number of chassis.

**Configuration Options:**

* **Input Options:** -400 (400 ±10% Volt Line to Line AC input, not available on 6000Lx, 12000Lx, 18000Lx), -480 (480 ±10% – 3 phase output only).
* **Output Options:** -AX (Auxiliary outputs, 26 VAC regulated and 5 VAC unregulated). Limits upper frequency to 800 Hz.
* **Multi-Chassis Configurations:** Models indicated by /2 or /3 in the model number denote the number of chassis (e.g., 9000Lx/2 for 9 kVA).

### Why Choose Aumictech for California Instruments 3000LX?

At Aumictech, we specialize in supplying high-end AC power source and analyzer equipment. Whether you need to simulate complex global AC power conditions, perform detailed power quality analysis, or ensure the compliance of your electronic equipment with stringent standards, our team ensures the California Instruments 3000LX delivers maximum performance and reliability for your application.

MPN

3000LX

Brand Name

California Instruments

You May Also Like