The Data Precision 8200 is a 6 1/2-digit DC voltage and current calibrator engineered for precision signal generation and circuit calibration. Microcomputer-controlled with Incrematic™ decade assist, this instrument delivers completely isolated analog outputs across multiple ranges with exceptional stability and settling performance. The 8200 suits both manual test benches and automated test equipment systems, particularly when configured with the optional IEEE-488 bus interface for remote programming and system integration.
Technical Specifications
DC Voltage Output
• Ranges: ±100 mV, ±10 V, ±100 V (±1 kV optional)
• Resolution: 0.1 µV (100 mV range), 10 µV (10 V range), 100 µV (100 V range)
• Basic Accuracy: 10 ppm of reading + 1 ppm of full scale
• Stability: <10 ppm over 60 days; <20 ppm over 6 months
• Temperature Coefficient: 1 ppm/°C of reading + 0.2 ppm of full scale (10 V range, 15–35°C); 2 ppm/°C of reading + 0.2 ppm of full scale (100 mV and 100 V ranges, 15–35°C)
• Settling Time: ≤1 ms (same range), ≤15 ms (range change)
• Output: Floating, 4-terminal with remote sensing (10 V and 100 V ranges); rear-panel shielded 2-terminal plus guard (100 mV range)
DC Current Output
• Range: 1 µA to 100 mA
• Basic Accuracy: 0.01% of reading + 1 µA
• Stability: 10 ppm of reading + 2 ppm of range
• Output: Floating, 2-terminal plus guard
– Key Features
• 6 1/2-digit display for high-resolution readout
• Microprocessor-assisted Incrematic™ control for rapid manual decade adjustment
• Completely isolated analog signals eliminate common-mode errors
• Fast settling eliminates range-change delays in test sequences
• Warm-up requirement: minimum 30 minutes to 1 hour
• Compact 3.5-inch high, 19-inch wide 19-inch rack-mount chassis
• Weight: under 17 pounds for field portability
• Universal power: 115 or 230 VAC with rear-panel configuration switch
– Typical Applications
• High-precision analog circuit characterization and acceptance testing
• Calibration of digital multimeters, precision voltmeters, and milliammeters
• Automated test equipment signal sourcing
• Instrument calibration laboratories
– Compatibility & Integration
Optional IEEE-488 (GPIB/HPIB) interface enables remote programming and system integration for ATE platforms, eliminating manual front-panel operation in automated workflows.