Call +1 ‪(484) 841-9341‬

NIST TRACEABILITY 

NIST-traceable calibration certificate available with data

Service Support

Service plans available for the products’ durability

FAST REPAIRS

1–3 day turnaround, 15-day return on repair policy

Software Updates 

Staying up-to-date with latest software updates.

Fluke 80E-10 Voltage Divider – 10 kV max.

SKU: 8.00E-09Categories: Calibration Equipment
Brand: Fluke
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The Fluke 80E-10 is a voltage divider designed to extend the measurement capability of multimeters to 10 kV. This allows standard multimeters to measure high voltage signals, such as those found in electrical power distribution systems and high-voltage equipment.

Request CalibrationRequest Repair
Product Starting From:

$450.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote 8.00E-09
Equipment info

The Fluke 80E-10 is a precision high-voltage divider that extends multimeter measurement capabilities to 10 kV DC. It converts high-voltage signals into proportional low-voltage outputs suitable for differential voltmeters and null-type potentiometers, making it essential for characterizing power distribution systems and high-voltage equipment. The instrument features precision wirewound resistors with low temperature coefficient and achieves 0.01% division ratio accuracy within the 18°C to 28°C operating band.

## Technical Specifications

• **Maximum Input Voltage:** 10 kV DC
• **Division Ratios:** Divide by 1,000 and divide by 10,000
• **Output Terminals:** 10 V and 1 V proportional outputs on binding posts (3/4″ centers)
• **Division Ratio Accuracy:** ±0.01% (18°C to 28°C); 2 ppm/°C outside this range (0°C to 50°C)
• **Division Ratio Stability:** ±0.01% per year after internal recalibration
• **Panel Meter Accuracy:** Approximately 3% of full scale
• **Operating Temperature Range:** 0°C to 50°C
• **Input Impedance:** Infinite impedance required for specified accuracy; 10 MΩ voltmeters introduce ~0.1% loading error on 10 V output
• **Input Connector:** MS3102A-18-165 (mating connector supplied)
• **Dimensions:** 7″ H × 8.5″ W × 8″ D
• **Weight:** Approximately 6 pounds

## Key Features

• Center-zero panel meter for magnitude and polarity indication
• Precision wirewound resistive elements with low temperature coefficient
• Separate grounding binding post on front panel
• Connector shell connected to chassis ground
• Meets MIL-T-945A vibration specifications
• Meets MIL-E-4970A shock requirements (half sine: 20 g, 11 milliseconds)

## Typical Applications

• Electrical power distribution system measurements
• High-voltage equipment characterization
• High-voltage signal precision measurement

## Compatibility & Integration

Designed for use with Fluke differential voltmeters or null-type potentiometers. Fluke differential voltmeters offer infinite input impedance at null, eliminating loading errors. Case suitable for both bench use and rack mounting.

MPN

8.00E-09

Voltage / Current Range

10 kV DC input; 10 V and 1 V proportional outputs