The HP/Agilent 85025E is a coaxial detector designed for accurate power measurements in the 10 MHz to 26.5 GHz frequency range. It features a type-N male connector and is ideal for various applications, including signal monitoring, power leveling, and automated testing.

Request CalibrationRequest Repair
Product Starting From:

$2,231.80

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote 85025E
Equipment info

The HP/Agilent 85025E is a coaxial detector engineered for precision power measurement across 10 MHz to 26.5 GHz. It detects both modulated (AC) and unmodulated (DC) microwave signals, delivering ±0.40 dB dynamic power accuracy at 10 GHz and absolute power accuracy of ±0.40 dB at 50 MHz when calibrated at 0 dBm. The detector integrates directly with HP 8757 Series scalar network analyzers for RF characterization, network analysis, and component validation in laboratory and production environments.

Technical Specifications

Frequency & Power
• Frequency range: 10 MHz to 26.5 GHz
• Nominal impedance: 50 Ω
• Maximum input power: +20 dBm (100 mW), ±10 VDC
• AC mode dynamic range: −16 dBm to −65 dBm
• DC mode dynamic range: +10 dBm to −60 dBm
• Return loss: Typically < 2.0 dB; warranted to 17 dB across frequency range

Measurement Performance
• Frequency response (warranted): ±0.5 dB from 10 MHz to 26.5 GHz in DC mode at −10 dBm input
• Dynamic power accuracy: ±0.40 dB at 10 GHz (power change < 30 dB post-calibration)
• Absolute power accuracy: ±0.40 dB at 50 MHz (calibrated at 0 dBm)

Physical
• Connector: 3.5 mm male (SMA-compatible)
• Form factor: Compact probe-style
• Weight: Approximately 100 g

– Key Features

• Dual-mode detection: AC and DC signal measurement
• Wide frequency coverage from 10 MHz to 26.5 GHz
• Tight return loss characteristics for minimal measurement error
• Field-proven accuracy specifications for repeatable results

– Typical Applications

• Scalar network analysis
• RF and microwave power measurement
• Telecommunications system testing
• Radar and aerospace component characterization
• Production calibration and validation

– Compatibility & Integration

Designed specifically for HP 8757 Series scalar network analyzers, including the HP 8757D. Not compatible with HP 8756 or HP 8755 analyzers.

MPN

85025E

Frequency Range

10 MHz to 26.5 GHz

Power Range (dBm)

+20 dBm to −65 dBm

Voltage / Current Range

±10 VDC

Brand Name

HP / Agilent