The Keysight E1416A is a single-slot, C-size VXI power meter. This module is designed for ATE systems and provides accurate power measurements from 10 MHz to 18 GHz. It features a wide dynamic range and fast measurement speeds, making it suitable for testing various RF and microwave components and systems in manufacturing and automated test environments.

Request CalibrationRequest Repair
Product Starting From:

$1,850.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote E1416A
Equipment info

The Keysight E1416A is a single-channel average power meter housed in a C-size, single-slot VXI module with message-based control for integrated ATE systems. It delivers fast, accurate power measurements across a broad dynamic range using the 8480 family of thermocouple and diode power sensors, plus the 11722A sensor. The instrument achieves ±0.02 dB or ±0.5% absolute accuracy with a 1.00 mW reference (50 MHz oscillator, ±0.7%, NIST-traceable) and operates from 100 kHz to 110 GHz depending on sensor selection.

Technical Specifications

Measurement Range & Accuracy
• Power range: −70 dBm to +44 dBm (100 pW to 25 W), sensor dependent
• Frequency range: 100 kHz to 110 GHz, sensor dependent
• Dynamic range: 50 dB in 10 dB ranges
• Accuracy: ±0.02 dB or ±0.5% (absolute mode); zero set ±0.5% of full scale on most sensitive range, ±1 count

Measurement Functions
• Power, offset (±99.99 dB), averaging, duty cycle
• Resolution: selectable 0.1, 0.01, 0.001 dB with auto filter mode
• Averaging: 1 to 1024 readings (powers of 2)
• Duty cycle: peak power display for rectangular pulse RMS conversions
• Result units: W, dBm (absolute), %, dB relative

Module Power & Thermal
• +5 V: 1.0 A (IPM), 0.1 A (IDM); +24 V/−24 V: 0.1 A / 0.01 A each
• Slot power: 12.20 Watts; air flow: 0.20 liter/s; pressure drop: 1.00 mm H2O

– Key Features

• Single-channel architecture optimized for automated test workflows
• Message-based VXI interface with VXIplug&play drivers
• Factory-calibrated reference with NIST traceability
• Selectable averaging and resolution for optimized measurement speed and precision
• Sensor offset capability for zeroing and relative measurements

– Typical Applications

Manufactured device characterization, power amplifier testing, RF component validation, and production acceptance testing in multi-instrument ATE racks.

– Compatibility & Integration

Works with 8480 series thermocouple and diode sensors plus the 11722A sensor. Connects via P1/2 VXI connectors in standard mainframes supporting message-based instrumentation.

MPN

E1416A

Frequency Range

100 kHz to 110 GHz

Power Range (dBm)

−70 dBm to +44 dBm

Voltage / Current Range

+5V @ 1.0A (IPM) / 0.1A (IDM); +24V @ 0.1A (IPM) / 0.01A (IDM); −24V @ 0.1A (IPM) / 0.01A (IDM)

Brand Name

Keysight / Agilent / HP