The LeCroy 9410 is a dual-channel digital oscilloscope delivering 150 MHz bandwidth and simultaneous multi-gigahertz sampling capability for transient and repetitive signal acquisition. Built on 8-bit flash ADC architecture with extendable 12-bit vertical resolution, the 9410 combines high-speed data capture with advanced triggering and waveform processing to support demanding design validation and signal analysis workflows.
Technical Specifications
Bandwidth & Sampling
• DC to 150 MHz (5 mV/div to 2.5 V/div at 50 Ω)
• DC to 125 MHz (1 mV/div to 5 mV/div at 50 Ω)
• DC to 125 MHz at probe tip (with 94XX-P01 probe)
• 100 MS/s transient sampling, simultaneously on both channels
• 4 GS/s repetitive signal sampling, simultaneously on both channels
Vertical Resolution & Sensitivity
• 8-bit flash ADCs per channel; extendable to 12 bits via averaging and filtering
• Vertical resolution: 0.4% full scale (8-bit) or 0.025% (12-bit with averaging/filtering)
• Calibrated sensitivity: 1 mV/div to 2.5 V/div continuously variable
• Effective sensitivity: 100 µV/division with vertical expansion and signal processing
Acquisition & Memory
• 10 kilowords per channel
• Segmentable into 2, 5, 10, 20, or 50 blocks
• Reference memories (C and D): 10K, 16-bit word capacity for waveform storage
Timebase
• Range: 2 ns/div to 1000 s/div
• Clock accuracy: ±0.01%
• Interpolator resolution: 5 ps
• Trigger-time accuracy: 20 ps RMS
Triggering & Signal Processing
• Smart Triggering: pre- and post-trigger, hold-off by time or event, delay by time or event
• TV trigger (PAL, SECAM, NTSC line/field)
• Logic trigger (three-input pattern)
• State-qualified and time/event-qualified triggering
• Glitch trigger with FASTGLITCH detection (< 5 ns)
• Waveform expansion: up to 200×
• Optional WP01 selective filtering (8–11 bit resolution)
• Optional WP02 Spectrum Analysis Firmware (1000-point FFT)
Interfaces & Power
• GPIB (IEEE 488) and RS-232-C remote control
• 110 or 220 V AC, 45–440 Hz
• 150 W consumption
• Probe power: ±15 V, +5 V DC
Environmental
• Operating: 0–50° C
• Rated: 5–40° C
• Humidity: < 80%
– Key Features
• Dual-channel simultaneous acquisition at up to 4 GS/s for repetitive signals
• Flexible triggering with glitch detection faster than 5 ns
• Extendable vertical resolution to 12 bits with signal averaging
• Segmented memory for capture of multiple events in a single acquisition
• CRT display for real-time signal visualization
– Typical Applications
Electronics design validation, transient event capture, repetitive signal analysis, debug workflows requiring high time resolution and memory depth.