The NI CAL-4353 is a front-mounting calibration accessory designed for use with compatible National Instruments devices. It provides a convenient and reliable method for calibrating equipment, ensuring accurate and dependable measurements. This accessory simplifies the calibration process, saving time and improving overall system performance. Its front-mounting design allows for easy access and integration into existing setups.

Request CalibrationRequest Repair
Product Starting From:

$0.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote CAL-4353
Equipment info

The NI CAL-4353 is a front-mounting calibration accessory engineered for the NI PXIe-4353 32-Channel, 24-Bit PXI Temperature Input Module. It enables precise calibration and verification of the module’s temperature measurement performance, facilitating straightforward integration into existing test setups.

Technical Specifications

Physical Dimensions & Weight
• 15.28 × 12.95 cm (6.02 × 5.10 in.)
• 525 g

Environmental Conditions
• Operating Temperature: 0 °C to 55 °C (IEC-60068-2-1, IEC-60068-2-2; MIL-PRF-28800F Class 3 low/Class 2 high)
• Storage Temperature: -40 °C to 71 °C (IEC-60068-2-1, IEC-60068-2-2; MIL-PRF-28800F Class 3)
• Operating Humidity: 10% to 90% noncondensing (IEC-60068-2-56)
• Storage Humidity: 5% to 95% noncondensing (IEC-60068-2-56)

Shock & Vibration
• Operating Shock: 30 g peak, half-sine, 11 ms pulse (IEC-60068-2-27; MIL-PRF-28800F Class 2)
• Operating Random Vibration: 5 Hz to 500 Hz, 0.3 grms
• Non-operating Random Vibration: 5 Hz to 500 Hz, 2.4 grms (exceeds MIL-PRF-28800F Class 3)

Electrical Protection
• Input to GND: ±30 VDC
• TC+ to TC–: ±30 VDC

– Key Features

• Compatible with 4-wire resistance and 2-wire voltage calibrators
• High-impedance source connection capability (Sense+/Sense–)
• 15-minute recommended warm-up time for measurement stability
• 1-year calibration interval

– Compatibility & Integration

The CAL-4353 connects directly to 4-wire resistance and 2-wire voltage calibrators. In certain configurations, TC+ and Sense+ or TC– and Sense– may be internally shorted; the accessory is not protected against faults between these pairs. High-impedance sources must connect only to Sense+/Sense– terminals.

Regulatory Compliance
Meets EN 61326-1 (Class A emissions; Basic immunity), EN 55011 (Group 1, Class A), and AS/NZS CISPR 11 (Group 1, Class A).