The Rohde & Schwarz ZN-Z32 Inline Calibration Unit Controller automates vector network analyzer (VNA) calibration across 10 MHz to 8.5 GHz, eliminating manual intervention in RF and microwave test environments. This CAN bus-controlled system manages multiple inline calibration modules simultaneously, delivering recalibration in seconds while maintaining consistent measurement uncertainty—critical for production automation, thermal vacuum (TVAC) chambers, and inaccessible test setups.
Technical Specifications
Frequency Range
• 10 MHz to 8.5 GHz
Controller System
• CAN bus communication interface
• R&S®ZN-Z30 controller unit manages calibration modules
• Supports up to 48 inline calibration units across two CAN bus lines (24 units per line)
• Scalable to virtually unlimited test ports with multiple controllers
• Supplied with two R&S®ZN-CAN2 CAN-bus control cables
Operating Environment
• Ambient temperature: +5 °C to +40 °C
Software
• R&S®ZN-Z3ASW application software for system configuration and calibration process management
– Key Features
• Automated periodic recalibration at push-button operation
• In-situ calibration—permanently connected to test setup, no disconnection required
• Multiport calibration handling, critical for phase-stable production testing
• TVAC chamber compatibility for thermally extreme environments
• Factory temperature characterization of inline calibration units included standard
• Recalibration time reduced to seconds for high port counts
– Typical Applications
• Automated production test environments
• Thermal vacuum (TVAC) testing with inaccessible measurement ports
• Multiport device characterization and verification
• High-frequency RF and microwave system calibration requiring frequent recertification
– Compatibility & Integration
Compatible Inline Calibration Modules
• R&S®ZN-Z32 (10 MHz to 8.5 GHz)
• R&S®ZN-Z33 (10 MHz to 40 GHz)
Compatible Vector Network Analyzers
• R&S ZNB
• R&S ZVA
• R&S ZVT