Call +1 ‪(484) 841-9341‬

NIST TRACEABILITY 

NIST-traceable calibration certificate available with data

Service Support

Service plans available for the products’ durability

FAST REPAIRS

1–3 day turnaround, 15-day return on repair policy

Software Updates 

Staying up-to-date with latest software updates.

Yokogawa MV1024 Portable Paperless Recorder, 24 Ch, 1 S (Fast sampling mode: 125 ms)

SKU: MV1024Categories: Data Loggers
Brand: Yokogawa
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The Yokogawa MV1024 is a portable paperless recorder with 24 channels and a fast sampling mode of 125 ms. It offers versatile data acquisition and recording capabilities in a compact and user-friendly design, ideal for various industrial and laboratory applications.

Request CalibrationRequest Repair
Product Starting From:

$3,480.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote MV1024
Equipment info

The Yokogawa MV1024 is a 24-channel portable paperless recorder for industrial and laboratory data acquisition. It captures thermocouple, RTD, and DC voltage signals with dual sampling modes: 1 second normal mode and 125 ms fast sampling mode. Data logs to internal flash memory (80 MB, 200 MB, or 400 MB) or external CompactFlash and USB storage. The 10.4-inch TFT color LCD (640 × 480 pixels) displays real-time measurements with configurable bar graphs and four viewing orientations. Ethernet connectivity enables remote monitoring via HTTP and FTP, with text and binary file export formats for post-acquisition analysis.

Technical Specifications

Input Channels and Signal Types
• 24 isolated input channels
• Thermocouple: up to 18 types
• RTD: up to 12 types
• DC voltage: ±50 V range with isolated inputs available

Sampling and Measurement
• Normal mode: 1 second interval with 16.7 ms, 20 ms, or 100 ms A/D integration time
• Fast sampling mode: 125 ms interval with 1.67 ms A/D integration time
• Trigger mode and free mode data recording
• Snapshot image capture to CompactFlash card

Memory and Storage
• Internal flash: 80 MB, 200 MB, or 400 MB (configurable)
• Example capacity: 70 days continuous recording at 1 second intervals across 12 channels with 200 MB memory
• External: CompactFlash cards, USB memory devices
• File formats: text (.txt) and binary

Display
• 10.4-inch TFT color LCD, 640 × 480 pixel resolution
• Display modes: vertical, horizontal, horizontal wide, separated horizontal
• Integral bar graph display
• Digital update rate: 1 second
• Tag labels: 16 characters maximum
• Message display: 32 characters maximum

Power and Thermal
• AC supply: 100 to 264 VAC auto-switching (operating range 90–132 VAC or 180–264 VAC)
• DC supply: 12 VDC or 24 VDC (operating range 10.0–28.8 VDC)
• Warm-up time: minimum 30 minutes

– Key Features

• Dual sampling modes minimize power supply noise while maintaining acquisition speed
• Isolated channel inputs for thermocouple and voltage reduce ground loop interference
• Ethernet (10BASE-T) with TCP, UDP, HTTP, FTP, and Modbus protocol support
• Web server displays real-time MV screen images via standard browser
• Portable form factor with multiple power supply options for field deployment

– Typical Applications

Temperature monitoring in manufacturing and process control, multi-channel sensor logging in laboratory experiments, long-duration data acquisition with minimal storage footprint, remote condition monitoring via Ethernet networks.

– Compatibility & Integration

Ethernet protocols: TCP, UDP, IP, ICMP, ARP, DHCP, HTTP, FTP, SMTP, SNTP, Modbus TCP, and proprietary MV protocol. Data export supports standard text format and encrypted binary files for integration with data analysis software.

MPN

MV1024

Voltage / Current Range

100-264 VAC (90-132 VAC or 180-264 VAC operating), 12 VDC or 24 VDC (10.0-28.8 VDC operating), Input: ±50 V DC

Brand Name

Yokogawa