The Advantest R3132 is a spectrum analyzer spanning 9 kHz to 3 GHz, purpose-built for wideband signal analysis and electromagnetic compliance testing. It delivers amplitude measurement capability from +30 dBm down to noise floor, with total level accuracy of ±1.5 dB maximum. The instrument features a 6.5-inch TFT color LCD display with VGA output support, enabling real-time signal visualization on external monitors. A built-in frequency counter provides 1 Hz to 1 kHz resolution with marker-based accuracy tied to frequency reference stability of ±2 × 10⁻⁶/year (±1 × 10⁻⁷/year with option). Resolution bandwidth ranges from 1 kHz to 3 MHz in 1-3-10 sequence, with optional narrow-band filters down to 30 Hz.
Technical Specifications
Frequency Performance
• Frequency range: 9 kHz to 3 GHz
• Frequency span accuracy: ≤ ±1%
• Frequency reference stability: ±2 × 10⁻⁶/year standard; ±1 × 10⁻⁷/year optional
• Counter resolution: 1 Hz to 1 kHz
• Residual FM: ≤ 60 Hz p-p over 0.1 s
Amplitude Characteristics
• Total level accuracy: ±1.5 dB maximum
• Measuring range: +30 dBm to average displayed noise level
• Maximum input level (preamplifier OFF): +30 dBm ±50 VDC
• Maximum input level (preamplifier ON): +13 dBm ±50 VDC
• Average noise floor (preamplifier OFF): −118 dBm + 2f(GHz) dB at 1 kHz RBW
• Average noise floor (preamplifier ON): −132 dBm + 3f(GHz) dB at 1 kHz RBW
• 2nd-order harmonic distortion: ≤ −80 dBc (mixer input −30 dBm, f ≥ 100 MHz)
• 3rd-order intermodulation distortion: ≤ −80 dBc (two-signal, −30 dBm mixer input, offset > 10 kHz)
• Residual response (preamplifier OFF): ≤ −100 dBm
• Residual response (preamplifier ON): ≤ −105 dBm
Resolution & Signal Purity
• RBW (3 dB): 1 kHz to 3 MHz standard; 30 Hz to 300 Hz optional
• Signal purity: ≤ −105 dBc/Hz at 20 kHz offset; ≤ −100 dBc/Hz at 10 kHz offset
– Key Features
• High-resolution 6.5-inch TFT color LCD with external VGA display output
• Integrated frequency counter with 1 Hz minimum resolution
• Floppy disk drive for trace data and measurement condition storage
• Built-in preamplifier option for extended dynamic range
• Selectable input attenuation and multiple RBW filters
• Auto-calibration for frequency response correction
– Compatibility & Integration
• GPIB interface for automated measurement control
• RS232 serial interface for system integration
• VGA output for networked display configurations