The Agilent/HP 8753ES with options 011 and UK6 is a 3 GHz vector network analyzer engineered for RF and microwave component characterization across R&D and production environments. Option 011 removes the internal test set, enabling integration with external test sets for flexible measurement configurations. This instrument delivers 110 dB dynamic range and supports dual independent measurement channels with four display channels.
Technical Specifications
• Frequency Range: 30 kHz to 3 GHz
• Dynamic Range: Up to 110 dB
• Measurement Channels: Two independent channels
• Display Channels: Four channels
• Frequency Resolution: 1 Hz
• Source Output Power: −5 to +20 dBm (with external test set, refer to test set specifications)
• Frequency Stability (Standard): Typically ±7.5 ppm (0° to 55° C); typically ±3 ppm/year
• Frequency Stability (Option 1D5): Typically ±0.05 ppm (0° to 55° C); typically ±0.5 ppm/year
• Magnitude Trace Noise (10 Hz IF BW): < 0.006 dB rms (30 kHz to 3 GHz)
• Magnitude Stability: 0.02 dB/°C (30 kHz to 3 GHz, typical)
• Phase Trace Noise (10 Hz IF BW): < 0.038° rms (30 kHz to 3 GHz)
• Phase Stability: 0.05°/°C (30 kHz to 3 GHz, typical)
• Standard Impedance: 50 Ω
• Test Port Connectors: 7-mm (50-ohm systems)
– Key Features
• Option 011 deletes internal test set for external test set integration (Agilent 85047A, 50-ohm or 85046B, 75-ohm compatible)
• Full two-port, adapter-removal, and TRL/LRM calibration methods supported
• HP-IB (GPIB) interface for remote operation and automated data transfer
• Large LCD color display with VGA output capability
– Typical Applications
Component characterization, filter and amplifier analysis, transmission line measurement, and interconnect validation in RF/microwave development and manufacturing.
– Compatibility & Integration
Designed for systems requiring external test set architecture. Option UK6 provides market-specific configuration. Option 075 enables 75 Ω operation with Type-N connectors.