The Agilent/Keysight 85110A is a pulsed RF S-parameter test set that extends network analyzer capabilities for characterizing devices under test in pulsed RF environments. Operating across 2 GHz to 20 GHz, this accessory delivers precise measurements required for microwave and RF component validation with repeatable accuracy in demanding laboratory settings.
Technical Specifications
Frequency Range: 2 GHz to 20 GHz
Measurement Type: Pulsed RF S-parameter characterization
Test Port Configuration: Selectable port 1 or port 2
Test Port Connectors: 3.5 mm male
RF Input Connector: 3.5 mm female (rear panel)
Impedance: 50 ohm nominal
DC Bias Supply: 500 mA, 40 Vdc maximum
Attenuation Range (Incident Signal): 0 to 90 dB in 10 dB steps
Nominal Operating Power Level: 0 to −3 dBm
Maximum Input Power (RF Input Connector): +14 dBm
Dynamic Range — Transmission Measurements:
2–8 GHz: Maximum +11 dBm (Port 2), Minimum −64 dBm (pulsed), −78 dBm (CW); Receiver dynamic range 75 dB (pulsed), 89 dB (CW); System dynamic range 64 dB (pulsed), 78 dB (CW).
8–18 GHz: Maximum +11 dBm (Port 2), Minimum −63 dBm (pulsed), −78 dBm (CW); Receiver dynamic range 74 dB (pulsed), 89 dB (CW); System dynamic range 62 dB (pulsed), 77 dB (CW).
18–20 GHz: Maximum +11 dBm (Port 2), Minimum −62 dBm (pulsed), −77 dBm (CW); Receiver dynamic range 73 dB (pulsed), 88 dB (CW); System dynamic range 60 dB (pulsed), 75 dB (CW).
– Key Features
• High power handling capability: measures devices with up to 20 W (+43 dBm) CW output when internal step attenuators and isolator are installed
• Input power protection: step attenuators provide +17 dBm threshold; exceeding this requires equivalent attenuation
• Connector torque limit: 90 N-cm maximum to protect test ports and connected adapters
– Compatibility & Integration
Designed for integration with the HP 8510 network analyzer platform.