The Amplifier Research PM2003 is a three-channel RF power meter engineered for precision measurements across RF and EMC applications. This instrument simultaneously displays two of three available channels on a menu-driven LCD display, delivering measurement speeds up to 200 readings per second in single-channel mode and 100 readings per second with dual channels. The PM2003 operates from 10 kHz to 40 GHz and measures power from −70 dBm to +44 dBm, with dynamic range up to 90 dB when paired with diode power heads and 50 dB with thermocouple heads. All specifications are dependent on the selected PH2000 Series power head.
Technical Specifications
• Frequency Range: 10 kHz to 40 GHz (sensor-dependent)
• Power Measurement Range: −70 dBm to +44 dBm (sensor-dependent)
• Dynamic Range: Up to 90 dB (diode heads); up to 50 dB (thermocouple heads)
• Measurement Speed: 200 readings/second (1 channel); 100 readings/second (2 channels)
• Power Reference Output: 50 MHz ±0.005%, −60 dBm to +20 dBm, 0.1 dB resolution steps
• Reference Output Accuracy (0–20°C, NIST traceable): ±0.055 dB at 0 dBm; ±0.075 dB (−39 to +20 dBm); ±0.105 dB (−60 to −40 dBm)
• Display: 20 character × 4 line LCD; simultaneous dual-channel readout with proportional bar graph
• Display Units: Absolute (watts, dBm) and Relative (dBr)
• Display Resolution: 5 digits (nW, μW, mW, W); 4 digits (dBm)
– Key Features
• Three independent measurement channels with simultaneous dual-channel viewing
• Source impedance 50 ± 1 Ω; SWR <1.05
• Harmonic output <−50 dBc
• IEEE-488 interface for remote control and data logging
• Recorder BNC output (0–10 V into 1 mW load; 9.09 kΩ source impedance)
– Typical Applications
• EMC testing and compliance verification
• RF communications system characterization
• Scientific research requiring high-speed, multi-channel power acquisition
– Compatibility & Integration
• Operates with PH2000 Series RF power sensors
• Emulates HP437, HP438, and Boonton 4230 command sets; supports SCPI
• Requires compatible power head and 5 ft. power head cable (supplied with each sensor ordered)