Call +1 ‪(484) 841-9341‬

NIST TRACEABILITY 

NIST-traceable calibration certificate available with data

Service Support

Service plans available for the products’ durability

FAST REPAIRS

1–3 day turnaround, 15-day return on repair policy

Software Updates 

Staying up-to-date with latest software updates.

Ando AQ6140 Multi-Wavelength Meter

SKU: AQ6140Categories: DWDM / CWDM Test Equipment
Brand: Ando
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The Ando AQ6140 is a high-performance optical spectrum analyzer designed for precise measurement and analysis of optical signals. It offers a wide wavelength range, high resolution, and excellent sensitivity, making it suitable for various applications in optical communication and research.

Request CalibrationRequest Repair
Product Starting From:

$3,950.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote AQ6140
Equipment info

The Ando AQ6140 is a multi-wavelength meter for precise optical signal measurement and analysis across Dense Wavelength Division Multiplexing (DWDM) systems. It simultaneously measures wavelength and power for up to 256 channels with high accuracy and resolution, displaying waveforms and results in real time on a 6.5-inch color LCD. The instrument supports grid monitoring aligned with ITU-T recommendations, drift and stability analysis, and includes marker functions for detailed wavelength and power comparisons.

Technical Specifications

• Wavelength Range: 1270 to 1650 nm (182 to 236 THz)
• Wavelength Absolute Accuracy: ±2 ppm (±0.003 nm at 1310 nm and 1550 nm)
• Minimum Resolvable Separation: 10 GHz or less (80 pm at 1550 nm)
• Maximum Simultaneous Channels: 256
• Maximum Input Level: +10 dBm (total of all lines)
• Safe Maximum Input Level: +18 dBm (total of all lines)
• Power Linearity: ±0.3 dB (input levels ≥ -30 dBm)
• Polarization Dependence: ±0.5 dB
• Power Accuracy: ±0.5 dB
• Display: 6.5-inch color LCD, 640 × 480 dots

– Key Features

• Simultaneous wavelength and power measurement for all channels with waveform display
• Grid monitoring function for wavelength shift detection and ITU-T compliance verification
• Time-series drift measurement and analysis for individual channels
• Stability analysis with 3D waveform display, time displacement tracking, and variable rate trap function
• Peak wavelength and power list generation
• 3D waveform and peak list visualization for temporal analysis
• Marker function for wavelength and power level difference analysis
• Internal memory stores up to 10 measurement configurations; floppy disk export supported
• Three-step operation maximum for standard functions

– Typical Applications

• DWDM system characterization and channel monitoring
• Optical transmission system evaluation
• Optical device performance analysis
• Light source stability and drift assessment

– Compatibility & Integration

• Fiber Type: 9/125 µm single-mode
• Connectors: FC, SC, ST, DIN, DIA
• Remote Interfaces: GPIB, RS-232
• Monitor Interface: VGA

MPN

AQ6140

Frequency Range

182 to 236 THz

Power Range (dBm)

+10 dBm to +18 dBm

Wavelength Range

1270 to 1650 nm

Brand Name

Ando

You May Also Like