The Anritsu MA2472D is a standard diode sensor engineered for precision power measurement across microwave frequencies. It delivers high dynamic range and exceptional accuracy for continuous wave (CW) and time division multiple access (TDMA) signals, making it ideal for GSM and GMSK-modulated systems. The sensor integrates with Anritsu ML2430, ML2480, and ML2490 series power meters through automatic calibration data retrieval, eliminating manual setup and enabling frequency-, power-, and temperature-compensated measurements.
Technical Specifications
Performance Characteristics
• Dynamic Range: 90 dB
• Sensor Linearity: Better than 1.8% below 18 GHz
• Rise Time: 4 µs (suitable for TDMA pulse detection)
• Temperature Accuracy: Less than 1.0% from 5°C to 50°C
Electrical
• Frequency Range: 10 MHz to 18 GHz
• Power Measurement Range: -70 dBm to +20 dBm (CW mode); -40 dBm to +20 dBm (Profile mode); -34 dBm to +20 dBm (Pulse/Mod mode)
• Standing Wave Ratio (SWR): <1.17 (10–150 MHz), <1.12 (150 MHz–2 GHz), <1.22 (2–12.4 GHz), <1.25 (12.4–18 GHz)
• RF Connector: N-type male
Physical
• Dimensions: 73 mm × 52 mm × 29 mm
• Weight: 2 lbs (shipping weight)
– Key Features
• Internal EEPROM calibration storage with automatic interpolation across frequency, power, and temperature
• Precision RF connectors with hexagon coupling nut for torque-wrench attachment
• Optimized 4 µs rise time for accurate TDMA burst characterization
• Wideband coverage from 10 MHz to 18 GHz with controlled impedance matching
– Typical Applications
• GSM and TDMA system power analysis
• GMSK-modulated signal measurement
• Broadband microwave power monitoring
• Multi-channel power meter integration
– Compatibility & Integration
Works with ML2430, ML2480, and ML2490 series power meters. ML2407A/08A and ML2437A/38A meters require firmware revision 3.10 or higher.