Call +1 ‪(484) 841-9341‬

NIST TRACEABILITY 

NIST-traceable calibration certificate available with data

Service Support

Service plans available for the products’ durability

FAST REPAIRS

1–3 day turnaround, 15-day return on repair policy

Software Updates 

Staying up-to-date with latest software updates.

DRIVER D122 UNIBLITZ VINCENT SHUTTER OPTICAL LASER OPTICS

SKU: D122Categories: Industrial Controllers & PLCs
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The Uniblitz D122 Shutter Driver is designed to operate Vincent Associates VS series shutters. It offers precise timing and control for optical and laser applications, ensuring reliable performance and synchronization. Ideal for research, microscopy, and industrial imaging.

Request CalibrationRequest Repair
Product Starting From:

$235.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote D122
Equipment info

The Uniblitz D122 Shutter Driver is a precision control unit engineered to operate Vincent Associates VS series shutters and all Uniblitz shutters at their maximum repetition rates. It delivers synchronized exposure control for optical and laser applications across research, microscopy, and industrial imaging systems. The driver accommodates both normally open and normally closed shutter configurations and features three selectable operational modes via key switch: Standard (STD) for external pulse or contact closure control, Line Interrupt (A) for AC line failure detection with manual reset capability, and DC Interrupt (B) for DC voltage monitoring with manual reset on continuity restoration.

Technical Specifications

Shutter Compatibility
• Operates all Uniblitz shutters including VS series
• Patented design enables shutters to reach maximum specified frequency

Operating Modes
• Standard (STD): Exposure controlled by external pulse source or switch contact closure
• Line Interrupt (A): Self-disables upon AC line failure; manual reset required
• DC Interrupt (B): Self-disables upon DC voltage level disruption at interrupt input; manual reset required

Input Control
• External pulse (BNC, TTL) or switch contact closure for exposure timing
• Pulsed on/off exposure control via external pulse or remote actuate cable
• Reset input accepts external pulse (BNC, TTL), front panel switch, or remote actuate cable; requires 25ms pulse minimum
• RS-232 serial interface via 910RS cable for all input control
• Rising or falling edge selection for all input functions

Output Signals
• Exposure pulse output: BNC, TTL, active-low, maximum +4VDC; duration determined by exposure setting or input pulse length
• Electronic synchronization output (Sync): BNC, TTL, active-low, maximum 0.2VDC when active
• +8VDC unregulated output for remote switching and control circuits

Electrical
• Input voltage: 115VAC (60Hz) or 230VAC (50Hz) auto-sensing
• Line fuse: 3AG 1/2 Amp (115VAC) or 1/4 Amp SB (230VAC)
• Shutter fuse: 3AG 1/2 Amp
• Standard three-prong power inlet

– Key Features

• Three operational modes selectable via key switch for enhanced flexibility
• Supports normally open and normally closed shutter configurations
• Multi-input control: external pulse, contact closure, RS-232 serial interface
• Edge-selectable input triggering (rising or falling)
• Daisy-chaining capability via exposure pulse output
• Auto-detection for dual AC line voltages

– Typical Applications

• High-speed microscopy and imaging systems
• Pulsed laser control and synchronization
• Research laboratory optical instrumentation
• Industrial vision and inspection systems

– Compatibility & Integration

Compatible with all Uniblitz shutters and Vincent Associates VS series. Interfaces via BNC connectors for pulse signals and TTL control, RS-232 serial link (910RS cable), remote actuate cables, and front panel controls.

MPN

D122

Voltage / Current Range

115VAC (60Hz) / 230VAC (50Hz); +4VDC max output; 0.2VDC max sync output; +8VDC unregulated output

Brand Name

Uniblitz/Vincent Associates

You May Also Like