The HP/Agilent 16092A Spring Clip Fixture is a two-terminal test accessory for precision impedance measurement of leaded components, axial and radial lead devices, and surface mount components (SMDs). The fixture combines interchangeable spring clip modules—one with parallel electrodes for SMD contact and another for lead component retention—enabling reliable, repeatable measurements across DC to 500 MHz. With APC-7 connectors, ±40V DC bias capability, and support for both open and short compensation, the 16092A integrates directly with industry-standard impedance analyzers and LCR meters for R&D, quality control, and production environments.
Technical Specifications
• Frequency Range: DC to 500 MHz
• Configuration: 2-terminal DUT connection
• DC Bias Range: −40V to +40V
• Terminal Connector: APC-7
• Electrical Length (typical): 0.34 cm
• Maximum Voltage: ±40V peak (AC + DC combined)
• Operating Temperature: 0°C to 55°C
• Physical Dimensions: 150 mm (W) × 70 mm (H) × 80 mm (D)
– Key Features
• Dual-module design: interchangeable SMD and leaded-component modules screw onto base plate
• Spring clip contacts accommodate specified component geometries
• Parallel electrodes for direct SMD contact
• Separate electrode pair for axial and radial lead components
• Standard open and short compensation with electrical length compensation
• SMD module: open achieved by electrode separation; short by electrode contact
– Typical Applications
• Component impedance characterization under bias conditions
• Quality assurance testing of passive and active components
• Production-line impedance verification
• Research and development measurements requiring wide frequency coverage
– Compatibility & Integration
Direct compatibility with Keysight/Agilent E4980A, E4981A, E4982A, E4991B, E5061B-3L5/3L4/3L3 (with options 005 + 16201A), E4990A (with 42942A and option E4990A-120), HP 4291A, Agilent 4191A, 4263B, 4268A, 4279A, 4284A, 4285A, 4288A, and 4294A (with 42942A).