The HP/Agilent 4352B is a high-performance signal analyzer designed for analyzing signals up to 3 GHz. It offers a wide range of measurement capabilities, including frequency, amplitude, and modulation analysis. Ideal for testing and troubleshooting communication systems, RF devices, and other electronic equipment operating in the microwave frequency range.

Request CalibrationRequest Repair
Product Starting From:

$2,069.80

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote 4352B
Equipment info

The HP/Agilent 4352B is a dual-mode signal analyzer for characterizing Voltage Controlled Oscillators (VCOs) and Phase-Locked Loops (PLLs) in RF wireless applications. Operating as both a signal analyzer and VCO tester, it covers 100 MHz to 3 GHz natively, with extension to 12.6 GHz via the optional HP 43521A downconverter. The instrument delivers laboratory-grade analysis for design validation alongside production-line measurement speed.

## Technical Specifications

**Frequency Measurement**
• Measurement range: 100 MHz to 3 GHz
• Frequency span: 2 MHz, 20 MHz, MAX
• Resolution: Measurement range ÷ 40,000 Hz
• Accuracy: ±(Measurement range × 0.1% + external timebase accuracy)
• Transient resolution: 50 Hz; time resolution: 12.5 µs

**RF Power Measurement**
• Input range: −10 to +20 dBm
• Absolute accuracy: ±3 dBm @ −10 dBm, RF ATT = 10 dB (typical)
• Relative accuracy: 0.5 dB (typical)
• Input impedance: 50 Ω; VSWR <1.5 (@ <2 GHz), <1.3 (typical)

**Phase Noise Measurement**
• Performance: −157 dBc/Hz @ 1 MHz offset (typical)
• Measurement points: Up to 801 from 100 Hz to 10 MHz offset
• Sweep time: 7.4 seconds
• Carrier lock: Automatic via carrier lock multi-mode PLL technology

**Spectrum Analysis**
• FFT spectrum analyzer capability included

**FM Deviation**
• Range: 0 to 200 kHz peak
• Accuracy: ±(2% + 0.1% of measurement) @ 1 kHz FM rate

**DC Control & Power Supply**
• VCO tuning supply: 0 to +15.5 V, 50 mA max.; 1 mV resolution; ±(0.2% + 2 mV) accuracy
• Control voltage source: 0 to +20 V, 100 µV steps, 20 mA max.; ±(0.1% + 20 mV) accuracy (Option 001: −15 to +35 V)
• Control voltage noise: <1 nV/√Hz @ 10 kHz offset

**Filtering & Signal Generation**
• HP filters: 50 Hz, 300 Hz
• LP filters: 3 kHz, 15 kHz, 20 kHz
• 1 kHz signal source for VCO characterization
• Residual FM: <3 Hz rms (300 Hz – 3 kHz bandwidth)

## Key Features

• Dual measurement modes: signal analyzer and VCO tester
• Excellent phase noise characterization with automatic carrier tracking
• Integrated DC supply and ultra-low-noise control voltage source for complete VCO tuning evaluation
• High frequency resolution and transient measurement capability
• GPIB interface for automated test sequences

## Typical Applications

• VCO characterization and tuning curve generation
• PLL phase noise and transient response evaluation
• RF component validation in wireless transceiver design
• Production-line acceptance testing of oscillators

## Compatibility & Integration

• Extends to 12.6 GHz with HP 43521A downconverter option
• DC control voltage source supports Option 001 extended polarity range (−15 to +35 V)
• GPIB control for integration into automated measurement systems

MPN

4352B

Frequency Range

100 MHz to 3 GHz (extends to 12.6 GHz with HP 43521A downconverter)

Power Range (dBm)

−10 to +20 dBm

Voltage / Current Range

VCO supply: 0 to +15.5 V @ 50 mA max.; Control voltage: 0 to +20 V @ 20 mA max. (Option 001: −15 to +35 V)

Brand Name