IFR Aeroflex ATC1200Y3 Transponder & DME Test Set

SKU: ATC1200Y3Categories: RF Test Accessories
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The IFR Aeroflex ATC-1200Y3 is a comprehensive test set designed for transponder and DME (Distance Measuring Equipment) testing. It is used to verify the functionality and performance of aircraft transponders and DME systems, ensuring they meet regulatory requirements and operate correctly. This test set is suitable for bench testing in maintenance facilities or ramp testing on aircraft.

Request CalibrationRequest Repair
Product Starting From:

$5,479.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote ATC1200Y3
Equipment info

The IFR Aeroflex ATC-1200Y3 is a dual-function test set for transponder and DME system validation in aircraft maintenance and flight-line operations. It performs Mode A and Mode C interrogation, transmitter power measurement, and complete DME ranging and velocity simulation with self-interrogate capability. The instrument generates shaped pulse trains meeting FAA reply requirements and supports TACAN signal simulation via 60 Hz amplitude modulation.

Technical Specifications

Transponder Testing
• Mode A and Mode C interrogation
• On-off ratio: >80 dB
• PRF generator: 50 to 5000 Hz variable
• Sync pulse to P1 leading edge: 2 to 5 µs
• P3 pulse output: nominal amplitude -15 V, width 0.5 to 3 µs
• Gating pulses for diode switch isolation: >80 dB on-off ratio

DME Testing
• Range measurement: -1 to 999.89 nautical miles in 0.01 nm steps
• Velocity measurement: 0 to 9990 knots in 10-knot steps
• On-off ratio: >85 dB
• Echo pulse pair injection at 30 miles from interrogation
• Echo amplitude adjustment: +1 dB to -10 dB relative to reply pulses
• TACAN simulation via 60 Hz modulation: 0 to 50% variable level
• All present and future DME channels supported

Signal Generation
• L-Band leveled generator with crystal control
• Frequency sweep: 950 to 1225 MHz continuous
• VOR paired channels: 108.0 to 139.95 MHz
• Direct frequency selection: 950 to 1225 MHz
• MHz readout in 1 MHz steps (Gz Mode)
• Transmitter frequency measurement via heterodyne method
• Frequency resolution: 500 kHz at 1090 MHz

Calibration Output
• Crystal-controlled calibrator
• Pulse spacing: 1.0 µs or 1.45 µs
• Amplitude: 1 Volt across 51 Ω load
• Phase adjustment for interrogation and reply pulse alignment

– Key Features

• Rapid transmitter frequency setting
• Self-interrogate function for R-NAV testing without external DME
• Suppressor pulse output in XPDR and DME modes
• Digital selection of distance and velocity
• Peak power and transmitter frequency measurement

– Typical Applications

• Bench testing in maintenance facilities
• Ramp testing on operational aircraft
• Transponder compliance verification
• DME system performance validation
• TACAN signal simulation and testing

– Compatibility & Integration

Designed for direct aircraft and test bench integration supporting both Mode A/C transponders and all DME channel configurations.

MPN

ATC1200Y3

Frequency Range

950 to 1225 MHz; VOR paired channels 108.0 to 139.95 MHz

Voltage / Current Range

1 Volt across 51 Ω load (calibrator output); P3 pulse nominal amplitude -15 V

Brand Name

IFR Aeroflex