The Keithley 2290-10 is a high-voltage DC power supply delivering up to 10 kV with maximum output current of 1 mA. Engineered for precise high-voltage sourcing and monitoring, it serves applications including device breakdown testing, insulation testing, high-voltage resistivity measurements, and high-energy physics research. The unit combines low output noise with selectable filters to enable accurate low-level current measurements. Built-in interlock controls disable high-voltage output when safety interlocks are disengaged, and an optional protection module safeguards connected low-voltage instrumentation from voltage spikes during device breakdown events.
## Technical Specifications
**Output Characteristics**
• Voltage Range: +100 V DC to +10,000 V DC (1 V resolution)
• Maximum Output Current: 1.000 mA DC (1 µA resolution)
• Output Ripple: 0.01% of full scale, VRMS (300 Hz–300 kHz); <1 VRMS maximum
• With selectable filters: 1.0 mVRMS @ 1 kV, 2.0 mVRMS @ 3 kV, 3.0 mVRMS @ 5 kV
**Accuracy & Regulation**
• Voltage Set Accuracy: ±0.06% of full scale or ±(0.01% of setting + 2.5 V)
• Voltage Display Accuracy: ±1 V typical; ±2 V maximum
• Current Set Accuracy: 0.5% of full scale or 0.01% ±(0.01% + 2.5 µA)
• Current Display Accuracy: ±1 µA typical; ±2 µA maximum
• Line Regulation: 0.001% for ±10% line voltage change
• Load Regulation: 0.04% for 100% load change (typical)
**Response & Stability**
• Rise Time (full load): <6 seconds to within 1 V of final value
• Settling Time (full load): <4 seconds to within 1 V of final value
• Discharge Time (full load): <1 second to under 10 V
• Recovery Time: 120 ms for 40% load step change (typical)
• Trip Response Time: <10 ms
• Voltage Stability: ±0.01% per hour; <0.03% per eight hours
• Temperature Drift: 50 ppm/°C (0° to 40°C, typical)
• Output Stored Charge: <20 µC maximum
• Current Limit Range: 0 to 105% of full scale
• Trip Current Range: 10 µA to 105% of full scale
## Key Features
• Selectable output filters for noise reduction in sensitive measurements
• Safety interlock system with optional protection module
• IEEE-488 (GPIB) and RS-232 serial control
• Analog control voltage input for external modulation
• Power Rating: 75 W / 10 W
## Typical Applications
• High-voltage device breakdown testing
• Component and insulation testing
• High-voltage resistivity measurements
• High-energy physics research
## Compatibility & Integration
Programming interfaces include IEEE-488 (GPIB) and RS-232, with analog control voltage input for system integration.