The Teledyne LeCroy WavePro 725Zi is a 4-channel benchtop digital oscilloscope built on the X-Stream II architecture for high-speed data acquisition and analysis. It delivers 2.5 GHz analog bandwidth at 50 Ω input with a 150 ps typical rise time, 40 GS/s maximum sample rate, and up to 256 Mpts of acquisition memory. The instrument features a 15.3″ widescreen color touchscreen display and supports multiple input impedance options: 50 Ω, 1 MΩ, and 10 MΩ with supplied probe.
Technical Specifications
• Analog Bandwidth: 2.5 GHz @ 50 Ω (-3 dB); 500 MHz typical @ 1 MΩ
• Rise Time: 150 ps (typical, 50 Ω)
• Analog Channels: 4
• Input Impedance: 50 Ω ±2%, 1 MΩ || 16 pF, 10 MΩ || 11 pF
• Maximum Input Voltage: ±5 Vrms @ 50 Ω; 250 V max @ 1 MΩ
• Vertical Resolution: 8 bits; up to 11 bits with Enhanced Resolution (ERES)
• Sensitivity: 50 Ω: 2 mV–1 V/div; 1 MΩ: 2 mV–10 V/div
• DC Gain Accuracy: ±1.5% of full scale
• Bandwidth Limiters: 20 MHz, 200 MHz
• Sample Rate: 40 GS/s maximum
• Memory Depth: 64 Mpts standard; up to 256 Mpts optional
• Time/Division: 20 ps/div–1000 s/div (real time); 20 ps/div–10 ns/div (RIS mode)
• Clock Accuracy: ≤1 ppm + 0.5 ppm/yr aging
• Jitter Noise Floor: 1 ps typical
• Channel-to-Channel Isolation: ≥100:1 @ 2 GHz; ≥40:1 @ 3 GHz; ≥20:1 @ 4 GHz
• Display: 15.3″ widescreen WXGA color touchscreen
• Operating System: Windows Vista Business 64-bit SP1 or Windows 7 64-bit
– Key Features
• X-Stream II architecture enables rapid waveform capture, zooming, and measurement
• Long acquisition memory supports extended time-domain analysis
• Multiple input coupling modes (AC, DC, GND) on 1 MΩ channel
• Dual input connector support: 50 Ω and 1 MΩ paths
– Typical Applications
• High-speed digital signal characterization
• Pulse and timing measurement
• Long-duration transient capture and analysis
– Compatibility & Integration
• ProBus input support on all channels
• Supplied high-impedance probe (10 MΩ || 11 pF)
• Internal common timebase across all four channels