The Marconi/IFR Aeroflex 6203B is a Microwave Test Set, covering a frequency range of 10MHz to 26.5GHz. This instrument is designed for comprehensive testing and analysis of microwave components and systems.

Request CalibrationRequest Repair
Product Starting From:

$0.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote 6203B
Equipment info

The Marconi IFR Aeroflex 6203B is a microwave test set engineered for scalar network analysis, power measurement, and frequency characterization across 10 MHz to 26.5 GHz. The instrument delivers real-time fault location with 0.1% accuracy and return loss measurement against distance, enabling rapid transmission line diagnostics. A synthesized signal source with 1 Hz resolution operates as either a CW generator or swept source, completing 400-point sweeps in under 200 ms with high stability and fast step times.

Technical Specifications

Frequency & Signal Generation
• Operating range: 10 MHz to 26.5 GHz
• Signal source: Synthesized, 1 Hz resolution
• Sweep modes: Start/Stop and Center/Span
• Sweep performance: 400 points in <200 ms
• Residual FM: <1.5 kHz
• Frequency stability: ±1 Hz ± frequency standard error (CW mode)
• Output power: –20 dBm typical (10 MHz–10 GHz), –15 dBm typical (10–20 GHz), –10 dBm typical (20–26.5 GHz); maximum +5 dBm, peak +27 dBm

Power Measurement
• Range: –70 dBm to +35 dBm (sensor-dependent)
• Dynamic range: 80 dB (AC detection), 70 dB (DC detection)
• Accuracy: ±1 dB (±0.5 dB typical, including flatness at 0 dBm)
• Linearity: <0.5 dB over guaranteed power range

Electrical Output
• Voltage/current output: Rear panel BNC, 1 V/GHz or 0.5 V/GHz selectable (20 V maximum)
• Voltage linearity: ±15 mV
• Input impedance: 50 Ω

Physical
• Dimensions: Approximately 24 × 18 × 16 inches
• Weight: 40 lbs

– Key Features

• Integral frequency counter and power meter
• Fault location with 0.1% accuracy
• Return loss (VSWR) and insertion loss measurement
• Hard keys with soft key functions and help messages
• GPIB port for HPGL-compatible plotter
• Parallel port for Epson FX series printer
• Optional 3.5-inch disk drive for results storage (Option 003)
• Field-replaceable RF connectors available (Option 002)

– Connectors

• RF output: MPC 3.5 mm female (Type 6203); optional 3.5 mm and N-type female (Option 002)
• Input port: Precision Type N female
• Fault location test port: Precision Type N female
• Return loss port: Precision Type N female
• Rear panel input socket: BNC

– Compatibility & Integration

Standard GPIB interface enables integration with automated test systems and data logging. Precision Type N connectors on measurement ports ensure compatibility with industry-standard transmission line and component test fixtures.

MPN

6203B

Frequency Range

10 MHz to 26.5 GHz

Power Range (dBm)

–70 dBm to +35 dBm

Voltage / Current Range

1 V/GHz or 0.5 V/GHz selectable (20 V maximum)