Call +1 ‪(484) 841-9341‬

NIST TRACEABILITY 

NIST-traceable calibration certificate available with data

Service Support

Service plans available for the products’ durability

FAST REPAIRS

1–3 day turnaround, 15-day return on repair policy

Software Updates 

Staying up-to-date with latest software updates.

National Instruments SCXI-1160

SKU: SCXI-1160Categories: Signal Conditioning Modules
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The National Instruments SCXI-1160 is a high-density, 16-channel isolated digital input module designed for use in SCXI (Signal Conditioning eXtensions for Instrumentation) systems. It provides reliable and robust digital input capabilities for industrial and laboratory applications, featuring channel-to-channel isolation to protect against ground loops and voltage transients.

Request CalibrationRequest Repair
Product Starting From:

$0.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote SCXI-1160
Equipment info

The National Instruments SCXI-1160 is a 16-channel general-purpose switching module built around electromechanical latching relays in SPDT (Form C) configuration. Housed in a single SCXI chassis slot, it routes and switches medium-level signals via 250 Vrms contact-to-coil isolation and low thermal offset. Control occurs through serial communication over the SCXIbus by a connected DAQ device or switch controller.

Technical Specifications

• Relay Configuration: 16 × SPDT (Form C) latching relays
• Switching Capacity: 2 A at 250 Vrms AC; 2 A at 30 VDC
• Maximum Switching Power: 60 W
• Contact-to-Coil Isolation: 250 Vrms
• Initial Contact Resistance: < 75 mΩ (typical)
• Thermal EMF: ≈ 3 µV
• Relay Operating Time: ≈ 10 ms
• Scan Rate: Up to 50 operations/sec
• Bandwidth: 1 MHz
• Voltage Category: CAT II
• Dimensions: 7.8 in. × 1.2 in. × 6.7 in. (1 chassis slot)
• Front Connector: 48-pin DIN C male
• Rear Connector: 50-pin ribbon header

– Key Features

• Each relay channel includes normally closed (NC), normally open (NO), and common (COM) terminals for flexible signal routing
• Low on-resistance and minimal thermal offset preserve signal integrity across switching operations
• Latching relay design eliminates continuous coil power consumption
• Parallel operation with SCXI-1160 or SCXI-1161 modules in the same chassis

– Typical Applications

Signal routing and switching in automated test systems, data acquisition setups, and industrial control environments where medium-voltage AC and DC switching is required.

– Compatibility & Integration

Integrates with National Instruments SCXI chassis systems and is controlled via the SCXIbus. Compatible with NI MIO-16, Lab-NB, Lab-PC, Lab-PC+, Lab-LC, PC-LPM-16, DIO-24, DIO-32F, DIO-96 boards, NI 4021 switch controller, and NI 4070/4060 DMMs. Supports NI-SWITCH driver, NI-DAQmx driver, LabVIEW, and LabWindows/CVI. Optional SCXI-1324 Terminal Block accessory available. UL, CE, CAN/CSA, IEC, C-Tick, and FCC Part 15 certified.

MPN

SCXI-1160

Frequency Range

DC to 1 MHz

Voltage / Current Range

2 A per channel, 250 Vrms AC / 30 VDC

Brand Name

National Instruments

You May Also Like