The Rohde & Schwarz ENV216 is a Two-Line V-Network (LISN) designed for conducted emission measurements. It provides a defined impedance for accurate and repeatable measurements of mains-borne interference within the frequency range of 9 kHz to 30 MHz. Featuring excellent RF characteristics and robust construction, the ENV216 is suitable for compliance testing according to CISPR standards.

Request CalibrationRequest Repair
Product Starting From:

$0.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote ENV216
Equipment info

The Rohde & Schwarz ENV216 Two-Line V-Network is a Line Impedance Stabilization Network (LISN) for conducted emission measurements on single-phase Equipment Under Test. It provides defined impedance across 9 kHz to 30 MHz to ensure accurate, repeatable disturbance voltage measurements in compliance with CISPR 16-1-2, EN 55016-1-2, ANSI C63.4, and MIL-STD-461D/E/F standards. The unit handles AC supply voltages to 240 V ±10 % and constant currents to 16 A, making it suitable for laboratory and on-site testing environments.

Technical Specifications

The ENV216 simulates impedance of (50 μH + 5 Ω) || 50 Ω with magnitude error limits of ±20 % and phase error of ±11.5° across its operating band. It accepts AC supply from 50 Hz to 60 Hz ±5 % or DC up to 50 V. For AC voltages below 90 V or DC below 50 V, an external plug-in power supply is required for logic operation. A built-in pulse limiter (switchable) and 10 dB attenuator pad maintain standard impedance presentation to the measuring receiver. A switch-selectable 150 kHz high-pass filter prevents receiver overdrive from low-frequency interference. The artificial hand (220 pF + 511 Ω) is supplied as standard.

Isolation between AC supply and measuring receiver port exceeds 0–40 dB from 9 kHz to 50 kHz (increasing linearly with frequency logarithm) and exceeds 40 dB from 50 kHz to 30 MHz.

– Key Features

• Simulated impedance (50 μH + 5 Ω) || 50 Ω with defined error limits
• Frequency range 9 kHz to 30 MHz
• Current handling up to 16 A constant
• Switchable pulse limiter and 150 kHz high-pass filter
• 10 dB attenuator pad for receiver protection
• Compact design and low weight for portable deployment
• Calibrated per CISPR 16-1-2 and ANSI C63.4

– Typical Applications

• Conducted emission testing per CISPR and EN standards
• Disturbance voltage measurement on single-phase loads
• EMC compliance validation
• Equipment certification testing
• On-site and laboratory measurements

– Compatibility & Integration

The ENV216 connects via female EUT connector with protective earth contacts (country-specific variants available) and RF output on N female (50 Ω). Remote control uses 25-contact D-Sub female with TTL-level inputs, compatible with Rohde & Schwarz measuring receivers.

MPN

ENV216

Frequency Range

9 kHz to 30 MHz

Voltage / Current Range

0–16 A constant, 0–240 V AC ±10 %, 0–50 V DC