Rohde and Schwarz ESCI EMI Test Receiver

SKU: ESCICategories: EMI / EMC Test Receivers
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The Rohde & Schwarz ESCI is an EMI test receiver used for EMC compliance testing. It measures electromagnetic interference in a frequency range from 9 kHz to 3 GHz. Features include a built-in preamplifier, various detectors (peak, average, RMS, quasi-peak), and automated test routines for efficient and accurate measurements.

Request CalibrationRequest Repair
Product Starting From:

$0.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote ESCI
Equipment info

The Rohde & Schwarz ESCI is a dual-function EMI test receiver and spectrum analyzer covering 9 kHz to 3 GHz, engineered for EMC compliance certification and product development. It delivers measurement uncertainty below 1 dB across its full frequency range, with a displayed average noise level of typically −155 dBm (RBW = 10 Hz, preamplifier on). The instrument combines an LSI frontend with digital signal processing to achieve high reliability and measurement accuracy across CISPR, EN, ETS, FCC, ANSI C63.4, VCCI, and VDE standards.

Technical Specifications

Frequency Coverage: 9 kHz to 3 GHz

Measurement Bandwidths:
• EMI measurement: 200 Hz, 9 kHz, 120 kHz, 1 MHz
• Resolution bandwidths (analyzer mode, −3 dB): 10 Hz to 3 MHz in 1/3/10 steps
• Digital filters: 1 Hz to 30 kHz
• Channel filters: 100 Hz to 5 MHz

Detector Suite:
• Peak (max/min), quasi-peak, average, CISPR average, RMS, RMS-average
• Up to three detectors simultaneously selectable
• CISPR 16-1-1 pulse weighting via quasi-peak detector

Sensitivity and Dynamic Range:
• Total measurement uncertainty: < 1 dB (to 3 GHz)
• Displayed Average Noise Level: typically −155 dBm (RBW = 10 Hz, preamplifier on)
• 1 dB compression point: +5 dBm nominal (without preselection)
• Phase noise at 500 MHz: typically −113 dBc (1 Hz) at 10 kHz offset

Preselection and Gain:
• Built-in 20 dB preamplifier
• 11 preselection filters with integrated switch selection
• ESCI7 model: twelve preselection filters extending to 7 GHz

Measurement Speed:
• Receiver mode: from 100 µs (SCAN mode)
• Spectrum analyzer sweep: 2.5 ms to 16,000 s (span ≥ 10 Hz)
• Zero span (time domain): 1 µs to 16,000 s resolution at 125 ns

– Key Features

• 21 cm TFT color display with simultaneous multiple trace measurement
• Programmable scan table with up to ten subranges
• Limit lines for civil standards compliance
• Frequency-dependent transducer factor management
• Time domain analysis for click interference measurement
• Fast adjacent-channel power (ACP) measurements

– Typical Applications

• EMC certification testing to CISPR, EN, FCC, and ANSI standards
• Pre-compliance product development and troubleshooting
• Spectral purity analysis and phase noise characterization
• Broadband and narrowband interference identification

– Compatibility & Integration

The ESCI complies with civil EMI standards including CISPR, EN, ETS, FCC, ANSI C63.4, VCCI, and VDE, enabling certification-grade measurements across regulated frequency bands.

MPN

ESCI

Frequency Range

9 kHz to 3 GHz

Power Range (dBm)

-155 dBm to +5 dBm

Brand Name

Rohde & Schwarz

You May Also Like