The Rohde & Schwarz ESPC10 is an EMC precertification test receiver designed for emission testing throughout product development and production stages. Built on full-compliance receiver architecture, it delivers precompliance EMI measurement capabilities that reduce risk and accelerate time-to-market before full compliance testing and CE marking. The instrument supports all major commercial EMI standards including CISPR, EN, ETS, FCC, ANSI C63.4, VCCI, and VDE.
Technical Specifications
Frequency Coverage
• Standard: 150 kHz to 1000 MHz
• Optional (ESPC-B2 module): 9 kHz to 1000 MHz
• Optional (retrofit to 2500 MHz): 1000 MHz to 2500 MHz
Synthesizer & Stability
• Frequency resolution: 10 Hz and 100 Hz modes
• Frequency error: < 3 × 10⁻⁶ after 30 minutes warm-up
Detection & Measurement
• Parallel detector channels: Average, Peak, and Quasi-Peak
• CISPR 16-1 compliant interference weighting down to 10 Hz pulse repetition frequency
• Automatic CISPR bandwidth selection across test frequency range
• Bandwidth complement: 2 fixed-tuned and 6 tracking filters (9 kHz–1000 MHz); 2 tracking filters (1000–2500 MHz)
• Measures voltage, field strength, current, and pulse spectral density with automatic transducer factor correction
RF Performance
• Integrated preselection for accurate narrowband measurements
• Attenuator: 0 to 70 dB in 10 dB steps
• Image frequency rejection: 70 dB (1st and 2nd IF)
• IF rejection: 70 dB
• Automatic overload detection across entire signal path (input and IF stages)
User Interface & Automation
• 8-digit LCD with switchable backlighting
• Windows-based EMI test software supplied standard
• Firmware macros for automatic and semi-automatic test routines
• FAST PRESCAN function for critical spectrum identification
• Automatic level calibration
• Limit-line comparison macros (EN standards and equivalent)
– Key Features
• Portable architecture supporting internal or external battery operation
• MIL-STD-810D shock qualification (40 g spectrum)
• MIL-T-28800D vibration qualification (Class 5)
• Operating range: −10 °C to 55 °C (no condensation); rated 5 °C to 45 °C
– Typical Applications
Precompliance EMI testing across consumer, industrial, medical, and telecommunications product development workflows.
– Compatibility & Integration
Supports CISPR, EN, ETS, FCC, ANSI C63.4, VCCI, and VDE standard protocols.