The Tektronix 7834 is a 400 MHz analog storage oscilloscope built around a multi-mode CRT storage architecture for capturing and analyzing high-speed transient waveforms. It accepts two vertical and two horizontal 7000-series plug-in modules, enabling customized measurement configurations across diverse applications. The instrument delivers exceptional writing speed through fast variable persistence mode—reaching 5500 Divisions/µs in reduced scan—making it ideally suited for single-shot event capture. Four distinct storage modes (bistable, variable persistence, fast bistable, and fast variable persistence) combined with normal and reduced-scan options provide the flexibility needed for both real-time and stored waveform analysis.
Technical Specifications
Bandwidth & Speed
• Bandwidth: Up to 400 MHz (with 7A19 or 7A29 vertical plug-ins); 300 MHz (with 7A24 vertical plug-in)
• Rise time: 0.9 ns (with 7A19, 7A29)
• Fastest calibrated sweep: 1 ns/Division
Writing Speed
• Fast Variable Persistence: 5500 Divisions/µs (Reduced Scan), 300 Divisions/µs (Full Scan)
• Fast Bistable: 776 Divisions/µs (Reduced Scan), 50 Divisions/µs (Full Scan)
• Variable Persistence: 12 Divisions/µs (Reduced Scan), 2 Divisions/µs (Full Scan)
• Bistable: 0.2 Divisions/µs (Reduced Scan), 0.03 Divisions/µs (Full Scan)
Display & Storage
• CRT: 8 × 10 divisions (0.9 cm/Division)
• CRT accelerating voltage: 10 kV or 12 kV (mode dependent)
• Storage modes: Bistable, Variable Persistence, Fast Bistable, Fast Variable Persistence
• Scan modes: Normal Scan, Reduced-Scan
Power & Environmental
• Input: 115 V or 230 V nominal, 50 to 400 Hz
• Maximum rated operating voltage: 250 V
• High-efficiency switch-mode power supply with voltage regulation
– Key Features
• Four 7000-series plug-in compartments for vertical and horizontal signal conditioning
• Z-axis input for intensity modulation
• Built-in calibrator for measurement verification
• Selectable reduced-scan mode to maximize writing speed for transient capture
– Typical Applications
• Single-shot waveform capture
• Transient analysis in high-speed electronics
• Complex signal characterization requiring stored reference traces
• Real-time signal monitoring with multiple measurement channels
– Compatibility & Integration
Fully compatible with Tektronix 7000-series vertical and horizontal plug-in modules. Optional configurations include EMC shielding, X-Y delay compensation networks, and elimination of readout and probe power connectors.