Tektronix TCP312 Probe, AC/DC Current, DC-100MHz, 30A DC

SKU: TCP312Categories: Sensors & Transducers
Brand: Tektronix
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The Tektronix TCP312 is an AC/DC current probe with a DC-100 MHz bandwidth and a maximum current of 30A DC. It is designed for use with Tektronix oscilloscopes and is ideal for measuring current in a variety of applications.

Request CalibrationRequest Repair
Product Starting From:

$3,480.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote TCP312
Equipment info

The Tektronix TCP312 is an AC/DC current probe engineered for precision current measurement on Tektronix oscilloscopes. It combines transformer and Hall effect technologies to deliver DC to 100 MHz bandwidth with 30 A DC maximum capability and 50 A peak current handling. The split-core design enables non-invasive insertion on conductors up to 3.8 mm, 5 mm, or 21 mm × 25 mm depending on configuration. Minimum measurable current reaches 1 mA, with 5 mA achievable at ±3% DC accuracy. Rise time of ≤3.5 ns or ≤23 ns supports fast transient capture across the full bandwidth.

Status indicators provide real-time visual feedback on probe operation and error conditions including degauss, probe open, overload, improper 50 Ω termination, and incompatible probe types. Low DC drift and noise characteristics enable reliable low-level current measurements. The probe operates across 0 °C to +50 °C at altitudes to 2000 m and carries 600 VRMS CAT I & II and 300 VRMS CAT III bare wire voltage ratings.

Technical Specifications

• Bandwidth: DC to 100 MHz
• Maximum DC Current: 30 A
• Maximum Peak Current: 50 A
• Minimum Current Measurement: 1 mA
• DC Accuracy Threshold: 5 mA at ±3% or 1 mA
• Rise Time: ≤3.5 ns or ≤23 ns
• Input Coupling: AC/DC
• Construction: Split-core
• Probe Cable Length: 1.5 m, 2 m, or 8 m
• Jaw Size (Maximum Conductor): 3.8 mm, 5 mm, or 21 mm × 25 mm
• Bare Wire Voltage Rating: 600 VRMS CAT I & II; 300 VRMS CAT III
• Probe Dimensions: 20 cm × 1.6 cm × 3.2 cm
• Probe Weight: 0.15 kg
• Operating Temperature: 0 °C to +50 °C
• Operating Humidity: 5 to 85% R.H.
• Operating Altitude: Up to 2000 m

– Key Features

• Low insertion impedance for minimal circuit loading
• Status indicator array signals degauss, overload, termination, and compatibility issues
• Low DC drift and noise floor for sub-5 mA measurements
• Maximum amp-second product: 50 A·μs to 1 A/V (amplifier-dependent)

– Compatibility & Integration

Operation requires the TCPA300 amplifier (90–264 V, 47–440 Hz, 50 W max, CAT II). Compatible with Tektronix oscilloscopes featuring TEKPROBE Level II or TekConnect interfaces. TPA-BNC adapter required for DPO3000, MDO/MSO/DPO4000, MSO/DPO5000, and DPO7000 series. Integrates with TDSPWR2 power measurement software.

MPN

TCP312

Frequency Range

DC to 100 MHz

Voltage / Current Range

30 A DC, 300 VRMS CAT III, 50 A peak; 600 VRMS CAT I & II

Brand Name

Tektronix