The Wavetek Model 184 is a 5 MHz sweep generator delivering six waveform types—sine, triangle, square, positive pulse, negative pulse, and DC—across a frequency range of 0.0001 Hz to 5 MHz. Engineered for precision testing and calibration, it combines broad frequency coverage with flexible sweep control and multiple output configurations. The instrument generates output amplitudes from 20 Vp-p open circuit down to 1 mV, with continuous DC offset adjustment between ±10V (open circuit) and ±5V into a 500 Ω load. An attenuator provides 0 to 80 dB control via 60 dB in 20 dB steps plus a 20 dB vernier.
Technical Specifications
• Frequency Range: 0.0001 Hz to 5 MHz
• Waveforms: Sine, triangle, square, positive pulse, negative pulse, DC output
• Output Amplitude: 20 Vp-p open circuit; 10 Vp-p into 50 Ω load; 1 mV minimum
• Attenuator: 0 to 80 dB (60 dB in 20 dB steps + 20 dB vernier)
• Output Impedance: 500 Ω
• DC Offset: ±10V (open circuit) or ±5V (into 500 Ω); peak + offset limited to ±10V or ±5V
• Symmetry Control: 1:19 to 19:1 (variable duty cycle and sawtooth ramps)
• Frequency Accuracy: ±2% of full scale (0.005 Hz to 5 MHz); ±4% of reading and ±2% of full scale (0.0005 Hz to 0.005 Hz)
• Square Wave Rise/Fall Time: Less than 30 ns into 500 Ω load
• Triangle Linearity: Greater than 99% (0.0005 Hz to 100 kHz)
• Harmonics: At least 30 dB down in the 1 MHz range
• Stability: ±0.05% short-term (10 minutes); ±0.25% long-term (24 hours)
• TTL Output: Typical rise/fall times 15 ns; drives up to 20 TTL loads
• GCV Output: 0 to +5V open circuit, 1 kΩ source impedance
• Trigger/Gate Input: 1 Vp-p to ±10V; 10 kΩ impedance, 33 pF; 50 µs minimum pulse width; 5 MHz max repetition rate
• Power: 90–110V, 105–125V, 180–220V, or 210–250V; 50–400 Hz; less than 25 watts
• Temperature: 0°C to +50°C
• Dimensions: 28.6 cm × 13.3 cm × 27.3 cm; 3.9 kg net
– Key Features
• Linear sweep with 1000:1 ratio across 10 ranges
• Adjustable sweep endpoints and sweep time (100 µs to 100 s)
• Triggered and gated outputs
• Proportional generator control voltage output
• Compact form factor: 11.25 × 5.25 × 10.75 inches
– Typical Applications
Component testing, filter characterization, frequency response measurement, circuit calibration, and signal conditioning verification in laboratory and field environments.
– Compatibility & Integration
Standard coaxial connectors for main signal output. Generator Control Voltage output enables integration with external frequency control systems and measurement instruments.