The Yokogawa FG310 is a synthesized function generator designed for precision signal generation across test and measurement applications. It produces standard waveforms—sine, square, triangle, and pulse—plus user-defined arbitrary waveforms with linear, step, and spline interpolation. The instrument delivers phase-continuous output, triggered burst generation, gated burst output, and DC voltage capability across a single output channel.
Technical Specifications
Waveform Generation
• Standard waveforms: sine, square, triangle, pulse
• Arbitrary waveforms: user-defined with interpolation options
• Sine frequency range: 1 µHz to 15 MHz
• Square frequency range: 1 µHz to 15 MHz
• Triangle, pulse, and arbitrary waveform frequency range: 1 µHz to 200 kHz
• Frequency resolution: 1 µHz or 9 digits maximum
• Output impedance: 50 Ω ±1%
Signal Purity
• Sine harmonic distortion at 100 kHz: 0.3% max. (rms, 2nd–5th harmonics)
• Maximum harmonic content: −55 dBc at 100 kHz, −45 dBc at 1 MHz, −35 dBc at 10 MHz, −25 dBc at 15 MHz
• Spurious: −55 dBc max. (1 kHz to 100 MHz range)
Amplitude and Offset Control
• Offset voltage range: ±10 V (1 mV resolution)
• Offset accuracy: ±(0.3% of setting + 0.5% of amplitude setting + 40 mV)
• DC output range: ±10 V (1 mV resolution)
• DC output accuracy: ±(0.3% of setting + 20 mV)
• Output attenuator: 1/1, 1/10, 1/100 with 0.2% accuracy
Modulation
• AM: 0% to 100% (0.01% resolution)
• FM deviation: 0 Hz to 7.5 MHz (1 µHz or 9 digits resolution)
• PM deviation: 0° to 360° (0.01° resolution)
• Offset modulation: 0 V to 10 V (1 mV resolution)
• PWM deviation: 0% to 50% (0.01% resolution)
• Supported modulation types: AM, FM, PM, DSB-AM, offset modulation, PWM
• Modulation sources: sine, triangle with variable symmetry, pulse with variable duty cycle, arbitrary waveform
Sweep Functions
• Patterns: linear, logarithmic, linear step, logarithmic step, arbitrary
• Sweepable parameters: frequency, phase, amplitude, offset voltage, duty cycle
• Simultaneous frequency and amplitude sweep capable
– Key Features
• Single output channel with phase-continuous operation
• Triggered and gated burst modes for synchronized waveform generation
• High-resolution frequency control enabling µHz-level precision
• Comprehensive modulation suite with multiple waveform sources
• Arbitrary waveform definition with scaled time and parameter axes
– Typical Applications
• Component and circuit characterization across audio to RF frequencies
• Signal conditioning and stimulus generation for sensor testing
• Modulation and sweep measurements in communications systems
• Arbitrary waveform synthesis for specialized test scenarios
– Compatibility & Integration
The FG310 operates with 50 Ω termination and accepts external trigger, gate, and modulation signals for integration into automated test systems.