The Keysight 10076C is a 100:1 passive probe engineered for high-voltage measurements up to 3.7 kV peak in demanding test environments. Operating across 500 MHz bandwidth with <0.7 ns risetime, it captures fast transients and high-frequency signal detail while maintaining measurement accuracy on power electronics and industrial circuits. The probe's 100 MΩ input impedance and ~3 pF capacitance minimize circuit loading. A 1.8 m cable with BNC connector and integrated wear indicator ensures safe, reliable deployment in field and lab settings.
Technical Specifications
• Attenuation Ratio: 100:1
• Bandwidth: 500 MHz (system bandwidth dependent on oscilloscope when used with ≤150 MHz instruments)
• Risetime (Calculated): <0.7 ns (with 500 MHz oscilloscope)
• Maximum Input Voltage: 3.7 kV peak (2.6 kV RMS); 4 kV CAT I (mains isolated); 1 kV CAT II (post receptacle mains)
• Input Resistance: 100 MΩ (1 MΩ oscilloscope input termination)
• Input Capacitance: ~3 pF
• Compensation Range: 6–19 pF
• Cable Length: 1.8 m
• Propagation Delay: ~8 ns
• Operating Temperature: 0 °C to 50 °C
• Humidity: ≤90% RH non-condensing
• Shock Rating: 400g 1/2 sine (tip); 50g 1/2 sine (BNC connector)
• Altitude: 3000 m (indoor use only)
• Pollution Degree: 2
– Key Features
• High-voltage isolation rated for mains-isolated measurements
• Probe readout capability for direct attenuation indication
• Built-in cable wear indicator—discontinue use if black contrasting color appears
• Rugged, lightweight construction for small component access
• BNC connector for secure oscilloscope coupling
– Typical Applications
• Power electronics testing
• High-voltage circuit characterization
• Transient and switching event capture
• Industrial equipment diagnostics
– Compatibility & Integration
Compatible with Keysight InfiniiVision and Infiniium oscilloscope series, plus other instruments accepting standard BNC probe connections.