The Keysight 85032B is a Type-N, 50 Ohm calibration kit engineered for precise network analyzer calibration across DC to 6 GHz. This kit delivers the open, short, and load standards required for establishing traceable measurement references on instruments with Type-N connectors.
Technical Specifications
Frequency Range: DC to 6 GHz
Connector Type: Type-N, 50 Ohms
Load Standards
• Return loss ≥ 49 dB (ρ ≤ 0.00355) DC to 2 GHz
• Return loss ≥ 46 dB (ρ ≤ 0.00501) above 2 GHz to 3 GHz
• Return loss ≥ 40 dB (ρ ≤ 0.01000) above 3 GHz to 6 GHz
Open Standards (Phase Linearity)
• Male: ±0.501° ±0.234 °/GHz
• Female: ±0.501° ±0.484 °/GHz
Short Standards (Phase Linearity)
• Male: ±0.441° ±0.444 °/GHz
• Female: ±0.490° ±0.385 °/GHz
Optional Type-N to 7-mm Adapters (Option 100)
• Return loss ≥ 30 dB (ρ ≤ 0.03162) DC to 6 GHz
• Four precision phase-matched adapters included
– Key Features
• Complete calibration triad: male and female opens, shorts, and matched 50 Ohm loads
• Pin depth specifications critical for electrical performance; verified using dedicated gaging processes
• Operating temperature range +15 °C to +35 °C with error correction within ±1 °C of calibration temperature
• Storage temperature range −40 °C to +75 °C
• Environmental rating: 0% to 95% relative humidity (non-condensing) at 40 °C
• Maximum operating altitude 4,500 meters
– Typical Applications
Network analyzer calibration for Type-N connectorized devices and systems from DC through microwave frequencies. Suitable for device characterization, transmission line measurements, and component testing requiring high-fidelity calibration standards.
– Compatibility & Integration
Designed for Agilent/Keysight network analyzers. Option 003 provides calibration to 3 GHz maximum for reduced-frequency applications. All calibration standards are traceable to Type-N mechanical specifications with documented pin depth verification.