Call +1 ‪(484) 841-9341‬

NIST TRACEABILITY 

NIST-traceable calibration certificate available with data

Service Support

Service plans available for the products’ durability

FAST REPAIRS

1–3 day turnaround, 15-day return on repair policy

Software Updates 

Staying up-to-date with latest software updates.

Agilent/Keysight 85520A

SKU: 85520ACategories: RF Calibration Equipment
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The Agilent/Keysight 85520A is a spectrum analyzer adapter designed to extend the functionality and compatibility of spectrum analyzers. It enables connection to various devices and facilitates signal analysis tasks. This adapter ensures accurate and reliable measurements, making it an essential tool for electronic testing and research applications.

Request CalibrationRequest Repair
Product Starting From:

$5,378.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote 85520A
Equipment info

The Agilent/Keysight 85520A is a four-standard mechanical calibration kit delivering Open, Short, Load, and Through (OSLT) references for RF network analyzer calibration. Operating across DC to 26.5 GHz with 3.5 mm male connectors and 50 Ω impedance, this compact module establishes traceable calibration foundations for precision S-parameter measurements.

Technical Specifications

Frequency Range: DC to 26.5 GHz

Connector Type: 3.5 mm male

Impedance: 50 Ω

Load Return Loss:
• 0 to 5 GHz: ≥ 42 dB
• 5 to 15 GHz: ≥ 36 dB
• 15 to 26.5 GHz: ≥ 32 dB

Open Phase Error:
• 0 to 5 GHz: ≤ 1.5°
• 5 to 15 GHz: ≤ 3.0°
• 15 to 26.5 GHz: ≤ 4.5°

Short Phase Error:
• 0 to 5 GHz: ≤ 1.0°
• 5 to 15 GHz: ≤ 2.5°
• 15 to 26.5 GHz: ≤ 4.0°

Through Standard:
• Insertion Loss (typical): ≤ 0.035 dB × √(f/GHz) for male-male connectors
• Electrical Delay: 115.888 ps for male-male connectors

Load DC Resistance: 50 Ω ± 0.5 Ω

Dimensions: 70 mm × 39 mm × 14 mm

Weight: 0.03 kg

Operating Temperature: +18 °C to +28 °C

Storage Temperature: -40 °C to +70 °C

– Key Features

• Full OSLT calibration suite in a single compact module
• Broadband coverage to 26.5 GHz with controlled phase and return loss
• Load standard maintains ≥ 32 dB return loss across the entire frequency range
• Through standard supports non-insertable device calibration (cables, filters, male-male assemblies)
• Tight mechanical tolerances: 50 Ω ± 0.5 Ω load resistance

– Typical Applications

• Network analyzer calibration and verification
• Transmission line and filter characterization
• Device S-parameter measurement setup
• Periodic calibration validation and standards traceability

– Compatibility & Integration

The 85520A integrates with RF instruments requiring 3.5 mm male connectors and 50 Ω system impedance. Connector torque specification is 8 lb-inch (90 N-cm). Protective caps are recommended during storage. Annual inspection intervals maintain measurement integrity.

MPN

85520A

Frequency Range

DC to 26.5 GHz

Brand Name

Agilent / Keysight

You May Also Like

Can't Find The Right Product? Contact Us.