The EADS/Racal/Dana 6060B is a synthesized signal generator offering precise frequency control and versatile modulation capabilities. Ideal for testing and calibrating communication systems, it delivers stable and accurate signals across a wide frequency range. It is equipped with various modulation modes, including AM, FM, and pulse modulation, suiting diverse application needs.

Request CalibrationRequest Repair
Product Starting From:

$1,100.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote 6060B
Equipment info

The EADS/Racal/Dana 6060B Synthesized Signal Generator delivers precise frequency control and multi-mode modulation for telecommunications testing, calibration, and RF system validation. Operating from 0.01 to 1050 MHz with 10 Hz resolution, the instrument generates signals from -127 dBm to +13 dBm (+19 dBm overrange) with 0.1 dB steps and ±1 dB guaranteed accuracy. The 6060B supports AM, FM, and pulse modulation modes, including low-rate FM down to 0.5 Hz bandwidth and fast-rise pulse modulation with approximately 15 nanosecond rise/fall times.

Technical Specifications

• Frequency Range: 0.01 to 1050 MHz
• Frequency Resolution: 10 Hz steps
• Amplitude Range: -127 dBm to +13 dBm (±1 dB accuracy)
• Amplitude Overrange: +19 dBm maximum
• Amplitude Resolution: 0.1 dB steps
• Non-Harmonic Spurious Products: <-60 dBc
• Harmonics: <-30 dBc
• Internal Reference: 10 MHz crystal oscillator with optional high-stability (±5×10⁻⁹/day) or medium-stability (±1×10⁻⁷/month) alternatives
• Power Consumption: 53 W total
• Operating Temperature: 0°C to 50°C
• Physical Form: C-size VXIbus module, 5.5 lb (2.5 kg)

– Key Features

• AM, FM, and pulse modulation modes with low-rate FM capability to 0.5 Hz
• 50 non-volatile memory slots for front-panel setup storage
• Built-in self-tests with numeric diagnostic codes
• Multi-trigger support via VXIbus TTLTRG, two external trigger lines, SCPI command, or internal programmable timer
• IEEE-488 (GPIB) interface available as Option 488
• SCPI-compatible command set with VXIplug&play driver support
• Dual voltage input options (87-106 VAC / 104-127 VAC or 191-233 VAC / 209-254 VAC)
• Rack-mountable with Y6001 Rack Mount Kit

– Typical Applications

• RF and microwave system testing
• Telecommunications equipment calibration
• Modulation performance verification
• Receiver sensitivity characterization

– Compatibility & Integration

SCPI-compatible messaging and VXIplug&play drivers enable straightforward integration into automated test systems. Optional IEEE-488 (GPIB) interface supports remote programming via talk/listen capabilities. Airflow requirement: 4.0 l/s @ 0.5 mm H₂O.

MPN

6060B

Frequency Range

0.01 to 1050 MHz

Power Range (dBm)

-127 dBm to +13 dBm (overrange to +19 dBm)

Voltage / Current Range

104-127 VAC, 191-233 VAC, 87-106 VAC, or 209-254 VAC