The Exfo NR5000 Radiometer Module is a portable optical power measurement instrument for precise irradiance and power readings across the ultraviolet and visible spectrum. It integrates with EXFO optical test platforms to deliver high-accuracy measurements in laboratory and manufacturing environments. The module employs a proprietary flat-response detector and non-imaging optical interface to minimize measurement variation caused by source radiance and intensity fluctuations, while supporting multiple measurement modes and auto-sensing capabilities.
Technical Specifications
• Wavelength Range: 250 nm to 600 nm
• Accuracy: ±5% over 250–600 nm range
• Power Range: 10 mW to 15 W
• Resolution: 10 mW
• Operating Voltage: 9 V (battery)
• Detector System: Proprietary flat-response optical detector; uniform sensitivity across 250–600 nm
• Optical Interface: Non-imaging design; minimizes beam profile dependence
– Key Features
• Measurement Modes: Power (Watts), irradiance (mW/cm² or W/cm²), real-time tracking, peak hold, relative (percentage of reference), and absolute (NIST-traceable) modes
• Light Guide Auto-Sensing: Detects 3 mm, 5 mm, and 8 mm industry-standard diameters automatically
• Smart Power Management: Auto-ON when optical signal ≥10 mW detected; auto-OFF after inactivity
• Auto-Ranging: Full-scale measurement range selected automatically
• Exchangeable Filters: Field-selectable wavelength ranges via EXFO filter cartridges (250–450 nm, 320–500 nm, 365 nm bandpass, 320–390 nm, 400–500 nm)
• Form Factor: Compact, hand-held design for portable operation
– Interfaces and Compatibility
• Optical Input: Proprietary connector; compatible with EXFO light guide adapters
• Adapter Support: Works with Lamp Output Adapter and Proximity Measurement Adapter for flood geometry measurements
• Light Guide Coupling: Requires EXFO-compatible light guide adapters for 3 mm, 5 mm, and 8 mm guides