The Fluke PM3084 is a 100 MHz, 4-channel analog oscilloscope engineered for signal analysis, waveform visualization, and electronics troubleshooting. It delivers 1% deflection factor accuracy across vertical and timebase measurements, supported by edge and TV triggering modes, delayed timebase, and AUTOSET/AUTOCAL functionality for rapid setup and performance optimization under variable environmental conditions.
Technical Specifications
Bandwidth & Sensitivity
• Bandwidth: 100 MHz
• Rise time: <3.50 ns
• Vertical sensitivity: 2 mV/div to 5 V/div (Channels 3 & 4: 2 mV/div to 12.5 V/div with continuous calibration)
• Input impedance: 1 MΩ ±1% || 25 pF ±2 pF
• Maximum input voltage: 400 V DC + peak AC (150 Vrms CAT III in 1 MΩ position)
• Input coupling: AC, DC
• Bandwidth limiter: 20 MHz -3 dB
Timebase
• Main and B sweep: 50 ns/div to 0.5 s/div in 1-2-5 sequence or continuous calibration to 1.25 s/div–20 ns/div
• Magnification: x10 (achieves 5 ns/div fastest sweep)
• Accuracy: 1%
Triggering & Display
• Modes: Auto free run, Triggered, Single
• Edge and TV triggering with hold-off up to 20 divisions
• 8 × 10 cm viewing area, 16.5 kV acceleration voltage
• Parallax-free graticule with variable illumination and on-screen settings readout
Display Modes
• CH1, (−)CH2, CH3, (−)CH4, and Add modes (CH1 & (−)CH2, CH3 & (−)CH4)
• ALTERNATING or CHOPPED mode (1 MHz)
– Key Features
• AUTOSET automatically optimizes channel, timebase, and trigger selection
• AUTOCAL provides fine calibration adjustment for enhanced accuracy
• Cursor-based voltmeter with extensive automated measurement capabilities
• Delayed timebase for detailed signal examination
– Interfaces & Integration
• RS-232 interface for full remote control (CPL protocol)
• IEEE 488.2 interface available as option (SCPI protocol)
• Optional factory-installed auxiliary Y-output from Channel 1 (BNC, 50 Ω, 10 mV/div in 50 Ω)
– Power
• Tapless switched-mode supply: 100–240 V AC, 50–400 Hz
• Line fuse: 1.6 AT delayed action, 250 V