The HP/Agilent 83487A is a dual-channel plug-in module for the 83480A, 54750A, and 86100A/B/C mainframes, delivering integrated optical and electrical characterization of high-speed communication waveforms. The optical channel operates across the 750–860 nm band with 2.85 GHz bandwidth and supports 62.5/125 µm multimode fiber, while the electrical channel provides user-selectable 12.4 GHz or 20 GHz bandwidth via 3.5 mm input connector. This two-slot module excels in research, development, and production testing—particularly for Fibre Channel and Gigabit Ethernet transmitter compliance verification.
Technical Specifications
Optical Channel
• Bandwidth: 2.85 GHz
• Wavelength Range: 750–860 nm
• Fiber Compatibility: 62.5/125 µm multimode (maximum 125 µm)
• Calibrated optical power measurement with User Cal wavelength accuracy
• Switchable datacom filters for 1063 Mb/s and 1250 Mb/s (Option 041)
• Reference receiver support: 0.155, 0.622, 1.066, 1.250 Gb/s
Electrical Channel
• Bandwidth: User-selectable 12.4 GHz or 20 GHz
• Input: 3.5 mm male connector
• General-purpose oscilloscope measurement capability
Display & Acquisition
• Resolution: 451 points (horizontal) × 256 points (vertical)
• Averaging: 1 to 4096 samples
• Graticule options: Full grid, axes with tic marks, frame only, or none
I/O & Controls
• External trigger input: ±1 V range
• Probe power connector
• Auxiliary power connector
– Key Features
• Dual optical–electrical measurement in single module occupying two mainframe slots
• Accurate reference receiver for Fibre Channel FC 1063 and Gigabit Ethernet compliance
• User Cal feature maintains calibrated measurements across the 750–860 nm range
• Flexible bandwidth selection optimizes signal fidelity or noise performance
• Integrated average power metering with switchable channel filters
– Typical Applications
• Fibre Channel transmitter compliance testing
• Gigabit Ethernet optical receiver evaluation
• Tributary signal analysis on high-speed links
• Optical waveform characterization in R&D and manufacturing
– Compatibility & Integration
Plug-in module for Agilent 83480A Digital Communications Analyzer, 54750A, and 86100A/B/C Infiniium DCA/DCA-J mainframes. Occupies two slots per mainframe. Compatible with 62.5/125 µm multimode fiber inputs and standard 3.5 mm electrical connectors.