The HP/Agilent 8483A is a thermocouple-based power sensor for average RF and microwave power measurement from 100 kHz to 2 GHz. Operating across a dynamic range of −30 dBm to +20 dBm, this 75-ohm sensor delivers accurate readings suitable for component characterization, system validation, and production testing. The Type-N male connector and compact form factor enable integration into benchtop and automated test systems. A mechanical adapter (HP 1250-0597) permits connection to 50-ohm power meter reference connectors; impedance mismatch correction requires calibration adjustment.
Technical Specifications
• Frequency Range: 100 kHz to 2 GHz
• Power Range: −30 dBm to +20 dBm
• Maximum Average Power: 300 mW
• Maximum Peak Power: 10 W
• Maximum Energy Per Pulse: 30 W·μs
• Maximum Input Power: +23 dBm
• Damage Threshold: +33 dBm
• Impedance: 75 ohms
• RF Connector: Type-N (male)
• Sensing Element: Thermocouple
• Dimensions: 30 mm W × 38 mm H × 105 mm L
• Weight: 0.2 kg
– Key Features
• Standing Wave Ratio: <1.80 (100 kHz–600 kHz); <1.18 (600 kHz–2 GHz)
• Operating Temperature: 0 °C to 55 °C
• Storage Temperature: −40 °C to +75 °C
• Operating Humidity: <95% RH
• Maximum Operating Altitude: 4,572 m
– Compatibility & Integration
The 8483A integrates with HP/Agilent power meter platforms: HP 435 Series, HP 436A, HP 437B, HP 438A, Agilent EPM Series (E4418A, E4419A, E4418B, E4419B), Agilent EPM-P Series (E4416A, E4417A), Agilent P-Series (N1911A, N1912A), Agilent E1416A VXI, and HP 70100A. Supplied 50-ohm to 75-ohm adapter requires impedance-mismatch compensation in 50-ohm reference systems.