The Keithley 246 is a high-voltage source measure unit (SMU) designed for precision sourcing and measurement of voltage and current. It combines a high-voltage power supply with a digital multimeter, offering capabilities for I-V characterization and testing of various electronic components and materials. It features a wide voltage range and precise control, making it suitable for applications requiring high-voltage sourcing and measurement.

Request CalibrationRequest Repair
Product Starting From:

$290.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote 246
Equipment info

The Keithley 246 High Voltage Supply delivers precision DC sourcing from 0 to 3100 V with maximum output current of 10 mA, integrated with built-in metering for voltage monitoring and polarity verification. This instrument combines high-voltage generation with accurate measurement capability, enabling I-V characterization across biasing, resistivity assessment, and leakage testing applications. Solid-state comparator amplifier and stacked series regulator tubes provide ±1% dial accuracy and exceptional long-term stability of ±0.01% + 2 mV over 8-hour periods following warm-up.

Technical Specifications

• Output Voltage: 0 to 3100 V DC; three calibrated in-line dials with 10 V steps; potentiometer interpolation (≤50 mV resolution); selectable positive or negative polarity relative to chassis
• Output Current: 10 mA maximum; electronic current limiting to <13 mA with automatic 1-second recovery
• Accuracy: ±1% of dial setting
• Stability: ±0.01% + 2 mV (after 1-hour warm-up, per 8-hour period)
• Line Regulation: 0.001% + 2 mV across 10% line voltage variation
• Load Regulation: 0.002% + 2 mV (no-load to full-load)
• Ripple and Noise: <1 mV RMS above 5 cps
• Power Input: 105-125 VAC or 210-250 VAC, switch-selectable, 50-60 Hz; 105 W consumption
• Output Connectors: MHV series (UG-931/U) on front and rear panels

– Key Features

• Overload-tolerant design withstands repeated overload conditions without degradation
• Metal-film resistors and selected zener diode ensure operational reliability
• Remote control and programming capability via rear-panel connectors
• Metering function provides real-time output voltage and polarity verification

– Typical Applications

• High-voltage biasing for photomultiplier tubes, ion chambers, photocells, and current detectors
• Voltage gradient studies and surface/volume resistivity measurements
• Capacitor leakage characterization

– Compatibility & Integration

Physical footprint: 5.25 in H × 19 in W × 10 in D; 25 lb net weight. MHV connectors enable integration into existing test systems requiring dual front/rear panel access.