The Keysight (Agilent) 467A is a dual-function power amplifier and regulated DC supply instrument for laboratory and production test environments. Operating from DC to 1 MHz, it delivers 10 W peak output power with selectable gain stages (X1, X2, X5, X10) or continuous variable gain from 0 to 10, making it suitable for signal conditioning, transducer excitation, and precision power sourcing applications. The unit combines low distortion amplification with stable DC output across five voltage ranges, all from a compact footprint drawing under 35 W at full load.
Technical Specifications
Amplifier Mode
• Frequency range: DC to 1 MHz
• Voltage gain (fixed): X1, X2, X5, X10
• Voltage gain (variable): 0 to 10 with 0.1% resolution
• Gain accuracy: ±0.3% DC–10 kHz; 1.0% 10–100 kHz; ±10% 100 kHz–1 MHz (>40 Ω load)
• Output: 10 W peak, >20 V peak, 0.5 A peak
• Distortion: <0.01% @ 1 kHz; <1% @ 100 kHz; <3% @ 1 MHz
• Input protection: 200 V p-p; short-circuit protected at all times
• Floating output: ±200 V DC above chassis ground
Power Supply Mode
• Output voltage ranges: ±1 V, ±2 V, ±4 V, ±10 V, ±20 V
• Output current: ±0.5 A peak; current limit <800 mA
• Voltage adjustment resolution: 0.1% of full output
• Ripple and noise: <5 mV p-p
• Temperature coefficient: <0.05%/°C or ±2 mV/°C (whichever is greater)
• Capacitive load stability: 0.01 µF or less
General Specifications
• Input impedance: 50 kΩ || 100 pF or 10 MΩ || <20 pF (selectable)
• Output impedance: 50 Ω
• Input-referred noise: <25 µV rms (1 MΩ source)
• Warm-up time: 30 minutes
• Power requirement: 115 or 230 V AC (±10%), 50–400 Hz; <35 W consumption
• Fuse: 0.8 A slow-blow (both voltage ranges)
– Key Features
• Three-terminal floating amplifier architecture
• Dual front-panel input/output banana terminals with rear BNC connectors
• Mode selection and gain controls with discrete and variable settings
• Integrated short-circuit and overvoltage protection
– Typical Applications
• Low-frequency signal amplification and impedance conversion
• Precision DC bias and excitation voltage sourcing
• Transducer and sensor signal conditioning
• Power device testing and characterization
– Compatibility & Integration
Banana and BNC terminations support standard lab cabling. AC power inlet accommodates 115/230 V mains with unified fuse protection.