The Keysight/Agilent/HP 436A is a general-purpose digital power meter designed for manual and automatic RF and microwave power measurements across the 8480 series of power sensors. It measures absolute and relative power with ±0.5% accuracy, displaying results in watts, dBm, or dB without manual conversion. The instrument combines intuitive front-panel operation with optional HP-IB programmability for automated test systems.
Technical Specifications
Measurement Range & Accuracy
• Frequency range: 100 kHz to 110 GHz (sensor dependent)
• Power range: −70 dBm to +44 dBm (approximately 0.1 nW to 25 W, sensor dependent)
• Power accuracy: ±0.5%
• Display formats: Watts, dBm, or dB (relative)
Calibration & Zeroing
• Calibration factor compensation for sensor-specific accuracy improvements
• Automatic zero capability with front-panel switch
• Zero set: ±0.5% of full scale on most sensitive range (typical)
• Zero carry over: ±0.2% of full scale when zeroed on most sensitive range
Reference Oscillator
• Internal 50 MHz oscillator with Type-N output
• Reference output power: 1.0 mW
• Single transistor design
Recorder Output
• Linear response proportional to indicated power
• 1 Volt at full scale, 0.316 Volts at −5 dB
• Output impedance: 1 kΩ
• BNC connector
Power Supply
• Power consumption: <20 VA (<23 VA with Option 022)
• Input: 100, 120, 220, or 240 V AC (±5%), 48–66 Hz and 360–440 Hz
– Key Features
• Peaking meter function for analog adjustments
• Optional HP-IB (IEEE 488) programmability (Option 022) with function codes AH1, C0, DC2, DTO, LEO, PO, RL2, SHI, SRO, T3, TEO
• Optional BCD interface control (Option 24)
• Removable power supply board with vertical board architecture
• Compact bench-top form factor: 134 mm H × 213 mm W × 279 mm D
• Net weight: 4.5 kg (10 lb)
– Compatibility & Integration
Full compatibility with the entire HP 8480 series power sensors, including thermocouple and diode types. Supported accessories include the HP/Agilent/Keysight 11730A Power Sensor cable (5 ft./1.5 m) and rackmount adapter kit (HP 5061-9657).