Call +1 ‪(484) 841-9341‬

NIST TRACEABILITY 

NIST-traceable calibration certificate available with data

Service Support

Service plans available for the products’ durability

FAST REPAIRS

1–3 day turnaround, 15-day return on repair policy

Software Updates 

Staying up-to-date with latest software updates.

Lambda LDS-X-15 DC Regulated Power Supply

SKU: LDS-X-15Categories: DC Power Supplies
Brand: Lambda
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The Lambda LDS-X-15 is a DC regulated power supply designed for providing stable and reliable DC power. It features a robust design and is suitable for a variety of applications requiring a dependable power source.

Request CalibrationRequest Repair
Product Starting From:

$125.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote LDS-X-15
Equipment info

The Lambda LDS-X-15 is a 15 VDC regulated power supply engineered to deliver stable, precision-regulated power in demanding technical environments. This convection-cooled unit combines tight voltage regulation, minimal output ripple, and automatic protection circuits to ensure reliable operation across temperature-sensitive and industrial applications. With selectable 105–127 VAC or 210–254 VAC input, the LDS-X-15 scales across a wide range of operating conditions while maintaining stringent electrical performance.

Technical Specifications

Output: 15 ± 5% VDC; maximum 5.5 A at 40°C ambient (4.7 A with cover), derating to 3.6 A at 60°C ambient.

Input Power: 200 W; 105–127 VAC or 210–254 VAC (transformer tap selectable); 47–440 Hz frequency range (ratings apply 57–63 Hz).

Line & Load Regulation: 0.005% + 0.5 mV across input variations (105–254 VAC) and load transitions (no load to full load).

Output Ripple & Noise: 150 µV RMS, 1 mV peak-to-peak (either polarity grounded).

Temperature Coefficient: ±(0.01% + 10 µV)/°C; operating range 0°C to 60°C continuous (with derating above 40°C); storage –55°C to +85°C.

Isolation & Safety: 10 MΩ minimum isolation (DC to ground, 750 VDC); MIL-T-27C Grade 6 transformer with 4000 VAC input/output isolation and electrostatic shielding.

– Key Features

• Zero overshoot on power-on, power-off, and failure transients
• Automatic electronic current limiting and thermal protection (auto-reset)
• Remote sensing provision eliminates lead-resistance voltage drop effects
• Fungus-inert construction; no nutrient materials used
• Heavy-duty PCB-mounted screw terminals
• Three mounting surfaces and mounting positions

– Typical Applications

Laboratory instrumentation, industrial control, telecommunications, and field-deployable systems requiring stable, regulated 15 VDC in temperature-variable or high-vibration environments.

– Compatibility & Integration

MIL-STD-810C validated for Low Pressure, High Temperature, Low Temperature, Temperature Shock, Temperature-Altitude, Humidity, Fungus, Vibration, and Shock. Physical dimensions 7¾” × 4⅞” × 2⅝”; net weight 7.3 lbs. Gray finish per Federal Standard 595, No. 26081.

MPN

LDS-X-15

Frequency Range

47–440 Hz (ratings apply 57–63 Hz)

Voltage / Current Range

15 ± 5% VDC; 5.5 A @ 40°C (4.7 A with cover), 3.6 A @ 60°C (3.1 A with cover), 4.6 A @ 50°C (3.9 A with cover)