The Philips PM2811 is a single-output programmable DC power supply delivering 0–30 V at 0–10 A (60 W) for precision bench-top and automated test applications. Built with GPIB/SCPI control, the PM2811 combines high output accuracy (±0.04% + 10 mV voltage; ±0.1% + 50 mA current) with fast programming response and comprehensive protection. Its autoranging architecture and internal memory support flexible voltage/current configurations within rated power limits.
Technical Specifications
Output Ratings
• Voltage: 0–30 V
• Current: 0–10 A
• Power: 60 W
• Single output module, 2U rack height
Accuracy & Stability (post warm-up, calibration conditions)
• Output voltage accuracy: ±0.04% of setting + 10 mV
• Output current accuracy: ±0.1% of setting + 50 mA
• Over voltage protection range: 2–32 V (±0.5% + 150 mV accuracy)
• Temperature coefficient: (0.016% + 2 mV)/K voltage; (0.05% + 2 mA)/K current
• Long-term drift (8-hour period): 0.04% + 5 mV voltage; 0.05% + 10 mA current
• Load regulation (10–90% change): ±(0.01% + 5 mV) constant voltage; ±(0.02% + 5 mA) constant current
Dynamic Response
• Programming speed: 100 ms @ 2 A load (0 V to max or max to 1 V); 50 ms to 1 V unloaded
• Reprogramming delay: 50–60 ms range, 1 ms resolution
• Current slew rate: <1 A/µs
• Settling band: ±50 mV; overshoot <150 mV (ΔI = 1 A)
Output Noise & Control
• Voltage ripple (PARD): 15 mV peak–peak / 3 mV rms (constant voltage)
• Current ripple: 10 mA rms
• Switching frequency: ~20 kHz
• Programming resolution: 12-bit (7.5 mV/LSB voltage; 2.5 mA/LSB current)
• Readback accuracy: ±(0.05% + 25 mV) voltage; ±(0.1% + 30 mA) current
– Key Features
• GPIB interface with SCPI command support for remote control and integration
• Over voltage protection with user-selectable trip threshold (2–32 V)
• Internal memory for storing output and protection settings
• Low output noise suitable for sensitive load testing
• Fast transient response with minimal overshoot
– Typical Applications
• Component characterization and burn-in testing
• Automated test system integration
• Laboratory bench-top power delivery
• Load simulation and equipment validation
– Compatibility & Integration
GPIB connectivity and SCPI command compatibility enable straightforward integration into automated test environments and remote operation via standard instrument control protocols.