The Stanford Research Systems FS710 is a 10 MHz AGC distribution amplifier that delivers seven stable outputs from a single input source. Built-in Automatic Gain Control compensates for up to 30 dB of cable loss, enabling 10 MHz reference distribution over distances approaching one mile. The transformer-coupled input accepts signals from 50 mVpp to 5 Vpp across the 10 MHz ± 100 kHz band, while all seven outputs deliver 1 Vpp into 50 Ω with ±10% stability. Output level is adjustable from 0.75 to 1.25 Vpp via front-panel potentiometer. Input and output impedance are jumper-selectable between 50 Ω and 75 Ω, with VSWR below 1.2 at 10 MHz on both ports. Harmonic distortion remains below -25 dBc across the output spectrum.
Technical Specifications
Input
• Frequency: 10 MHz ± 100 kHz
• Type: Insulated BNC, transformer coupled
• Level: 50 mVpp to 5 Vpp
• VSWR: <1.2 at 10 MHz
• Impedance: 50 or 75 Ω (jumper selectable)
Output
• Quantity: 7 local grounded BNCs
• Level: 1 Vpp into 50 Ω, ±10% (adjustable 0.75–1.25 Vpp)
• VSWR: <1.2 at 10 MHz
• Distortion: < -25 dBc
Gain Control
• AGC range: Up to 30 dB compensation
Power
• Input: 100/120/220/240 VAC, 50/60 Hz
• Consumption: 3 W
Physical
• Dimensions: 7.75" W × 2" H × 7.5" D
• Weight: 3 lbs
Environmental
• Operating temperature: 0°C to 40°C
– Key Features
• Single input drives seven independent outputs with AGC stabilization
• Transformer-coupled input isolation
• Dual impedance selection for flexible system integration
• Potentiometer-adjustable output amplitude
• Low harmonic distortion for precision timing applications
• Compact rack-compatible form factor
– Typical Applications
• Multi-instrument 10 MHz clock distribution in measurement systems
• Long-distance frequency standard distribution with automatic cable loss compensation
• Laboratory timebase routing to geographically dispersed equipment
– Compatibility & Integration
The FS710 accepts standard BNC connectors on input and output. Impedance selection via jumper allows operation in both 50 Ω and 75 Ω systems. Universal AC mains voltage support (100–240 VAC) accommodates global installation environments.