The Wiltron 560A is a GPIB-compatible scalar network analyzer for amplitude measurements of microwave components and systems. This three-channel instrument simultaneously measures transmission loss/gain, return loss, absolute power, and SWR across a broad frequency range determined by your choice of SWR Autotester and detector. Two independent measurement channels (A and B) plus a reference channel deliver dual simultaneous measurement display, making it ideal for rapid characterization of RF devices including filters, amplifiers, and passive networks.
Technical Specifications
• Frequency Range: 10 MHz to 40 GHz (dependent on selected SWR Autotester and detector)
• Measurement Channels: Two measurement channels (A and B) plus one reference channel (R); simultaneous dual-channel display capability
• Dynamic Range:
– Channels A and B: Up to 71 dB (+16 dBm to −55 dBm), usable to −65 dBm
– Channel R: 46 dB (+16 dBm to −30 dBm)
• Measurement Parameters: Transmission loss/gain, return loss, absolute power, SWR, voltage (with optional Volt Mode Adapter Cable)
• Return Loss Accuracy: Determined by SWR Autotester directivity — 40 dB to 18 GHz, 32 dB to 26.5 GHz, 30 dB to 40 GHz
• Detector Return Loss: 22 dB (0.04–8 GHz), 17 dB (to 20 GHz), 13 dB (to 26.5 GHz), 12 dB (to 40 GHz)
• Display: X-Y plot mode with Tektronix 620 CRT monitor integration
• Normalization Memory: Stores transmission residuals and return loss calibration data with automatic system residual subtraction
– Key Features
• Zero-biased Schottky diode detectors minimize drift and circuit complexity
• Three independent measurement channels for flexible test configurations
• Automatic calibration residual subtraction during normalization
• Non-flickering display refresh memory
– Typical Applications
• RF filter characterization
• Amplifier gain and loss measurements
• Component return loss and SWR testing
• Broadband passive network analysis
– Compatibility & Integration
• GPIB programmability with Option 3 for automated measurements
• Compatible with Wiltron 85 and HP controller systems
• Portable and rack-mount configurations available