The Agilent/Keysight 3468A Option 001 is a 5½ to 3½ digit digital multimeter delivering six measurement functions: DC voltage, true RMS AC voltage, 2-wire resistance, 4-wire resistance, DC current, and true RMS AC current. Display resolution is selectable across 5½, 4½, or 3½ digits, trading measurement speed against noise rejection as application demands.
Autoranging operates across all functions. The instrument features closed-box electronic calibration—eliminating internal adjustments and executable manually or remotely via the built-in HP-IL interface. Remote programmability provides control over functions, ranges, and measurement modifiers. Option 001 includes a rechargeable battery pack for field operation.
Technical Specifications
DC Voltage: 30 mV to 300 V with 1 µV sensitivity. Input resistance exceeds 10 GΩ on ranges to 3 V; 10 MΩ ±1% on 30 V and 300 V ranges. Maximum non-destructive input: 300 Vrms or 450 V peak. Accuracy specified as ±(0.0035% of reading + 2 counts) on the 3 V range over a 24-hour period.
AC Voltage (True RMS): 0.3 V to 300 V, 1 µV sensitivity. Bandwidth extends 20 Hz to 100 kHz on all ranges; 20 Hz to 300 kHz on the 30 V range. Crest factor >4:1 at full scale. Input impedance: 1 MΩ ±1% in parallel with 10% of full scale.
Resistance (2-wire/4-wire): 30 Ω to 30 MΩ with 1 mΩ sensitivity. Open circuit voltage <6.5 V. Test current varies: 1 mA (300 Ω, 3 kΩ), 1 µA (300 kΩ, 3 MΩ), 100 nA (30 MΩ). 2-wire accuracy matches 4-wire plus maximum 100 mΩ offset.
DC Current: 300 µA to 3 A, 10 µA sensitivity. Maximum non-destructive input: 3 A from <250 V source; fuse protected.
AC Current (True RMS): 300 µA to 3 A.
General: Operates 0 to 55°C, 95% RH (0 to 40°C). AC line powered 48 to 440 Hz, 86 to 250 V. Dimensions: 98.4 mm H × 238.1 mm W × 276.2 mm D. Weight: 2.1 kg.
– Key Features
• Selectable digit resolution for speed/noise optimization
• Closed-box calibration eliminates drift-prone adjustments
• HP-IL interface for remote integration into test systems
• Rechargeable battery option for portable environments
• True RMS AC measurements to 300 kHz
• 4-wire resistance measurement eliminates lead resistance error
– Compatibility & Integration
HP-IL connectivity enables integration into Hewlett-Packard automated test systems and compatible instrumentation networks.