The Agilent/Keysight 34923A is a 40/80-channel Reed Multiplexer module designed for use with the 34980A Multifunction Switch/Measure Unit. It offers high-density switching with excellent signal integrity, suitable for automated testing and data acquisition systems.

Request CalibrationRequest Repair
Product Starting From:

$885.58

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote 34923A
Equipment info

The Keysight 34923A is a 40/80-channel Reed Multiplexer module for the 34980A Multifunction Switch/Measure Unit mainframe. It delivers high-speed switching for automated test applications and data acquisition tasks, supporting flexible channel configurations: 40 two-wire channels, 80 one-wire channels, or mixed 2-wire and 4-wire arrangements on a single module. Reed relays enable scan rates up to 500 channels per second when connected to the mainframe’s internal DMM.

Technical Specifications

• Channel Configuration: 40 two-wire, 80 one-wire, or mixed 2-wire and 4-wire on the same module
• Switching Speed: Up to 500 channels/second (with 34980A internal DMM)
• Relay Type: Non-latching reed switches organized in two equal banks
• Switching Voltage: ±150 V peak
• Switching Current: 0.5 A
• Carry Current: 1.5 A
• Input Protection: 100 Ω protection resistors on each relay
• Direct Current Measurement: Channels 41–44 only; external shunts required for all other channels
• Analog Bus Relays: Eight armature relays (four per bank) for Analog Bus connection
• Relay Counter: Built-in counter for maintenance planning

– Key Features

• Flexible mixed-mode configurations support diverse test requirements
• Non-latching relay design with integrated input protection
• Standard 50-pin D-sub connectors compatible with existing infrastructure
• Optional terminal blocks available for 2-wire (34923T-001) and 1-wire (34923T-002) configurations
• Relay counter enables predictive maintenance scheduling

– Typical Applications

• High-throughput automated test systems
• Multi-channel data acquisition and monitoring
• General-purpose scanning with rapid channel switching
• Current measurement via user-supplied external shunts

– Compatibility & Integration

Designed exclusively for the Keysight 34980A mainframe. Up to eight plug-in modules can be installed in a single mainframe. Multiple multiplexer modules connect to the built-in Analog Buses, enabling expanded scanning capacity across the system.